Abstract

A Fresnel zone plate (FZP) with 35 nm outermost zone width has been fabricated and tested in the hard X-ray region. The FZP was made by electron beam lithography and reactive ion etching technique. The performance test of the FZP was carried out by measuring the focused beam profile for coherent hard X-ray beam at the beamline 20XU of SPring-8. The full width at half maximum of the focused beam profile measured by knife-edge scan method is 3 4 . 9 ± 2 . 7  nm, that agrees well with the theoretical value of diffraction-limited resolution. Applications to scanning microscopy were also carried out.

1. Introduction

Many types of optical elements for X-ray microfocusing and microimaging in the hard X-ray regions have been developed. Fresnel zone plates [1], total reflection mirrors [2], multilayer mirrors [3], refractive lenses [4], and multilayer Laue lenses [5] were fabricated and applied to microfocusing of synchrotron radiation X-rays. At present, the Fresnel zone plate (FZP) manufactured by electron beam lithography is one of the best optical elements for X-ray microscopy because of its high spatial resolution (better than 50 nm), negligibly small coma and small field distortion. The chromatic aberrations of FZP are usually ignorable in the hard X-ray region, because the crystal monochromators used in the synchrotron radiation beamlines produce sufficient monochromaticity for hard X-rays ( 𝜆 / Δ 𝜆 1 0 0 0 0 ). Thus, the spatial resolution of FZP optics is simply determined by the numerical aperture and by the accuracy of zone structure.

Spatial resolution of about 30 nm is already achieved with conventional FZPs. However, these results are carried out using third-order diffraction of FZP with an outermost zone width of 50 nm [1, 6]. Although a gold zone plate with 24 nm outermost zone width is also fabricated and tested in hard X-ray region [7], its spatial resolution has not been reported. Some X-ray focusing optics of spatial resolution better than 30 nm is already reported [2, 5]. However, they are still one-dimensional focusing, and these optical elements, the multilayer Laue lenses, or the aspherical Kirkpatrick-Baez mirrors are not image-forming optics. The Fresnel zone plate in the hard X-ray region is an almost ideal image-forming optical devise up to a spatial resolution of around 10 nm [8, 9]. Therefore, the conventional FZP is still indispensable for hard X-ray microscopy. In this report, recent development of FZP with an outermost zone width of 35 nm is presented, and characterization of the FZP in the hard X-ray region is described.

2. Fresnel Zone Plate Fabrication

The FZP was fabricated at NTT-AT Nanofabrication by electron beam lithography and reactive ion etching technique. The fabrication process of FZP is shown in Figure 1. The substrate is composed of 2  𝜇 m silicon nitride (SiN) membrane deposited on a Si wafer by low pressure chemical vapor deposition (CVD) method. Tantalum film with a thickness of 0.175  𝜇 m was deposited by rf-sputtering method, and SiO2 film was deposited by electron cyclotron resonance (ECR) plasma CVD method on the Ta layer as an etching mask of Ta. The FZP patterns were written by point electron-beam exposure method. The SiO2 film was etched with resist mask by the reactive ion etching, and the Ta film was dry-etched with SiO2 mask by an ECR ion-stream etching. Finally, the Si wafer was back etched with KOH solution for windowing. The details of fabrication method of tantalum zone plates at the NTT-AT Nanofabrication are already reported elsewhere [10, 11].

Schematic drawing of the FZP structure and its SEM image is shown in Figure 2. The design width of outermost zone that determines the diffraction-limited resolution is 35 nm. The SEM image shows a cross section of outermost zone structure that was prepared from the duplicated FZP. Fine zone structures of 35 nm width are precisely formed as shown in the figure. The zone pattern is designed by the simple formula of FZP as 𝑟 2 𝑛 = 𝑛 𝜆 𝑓 , ( 1 ) where 𝑛 is integer of zone number, 𝜆 is an X-ray wave length, and 𝑓 is focal length. We used a parameter set of 𝜆 = 0.155 nm ( 𝜈 = 8 keV) and f = 40 mm. Comparing with the spherical-aberration-corrected exact equation of zone pattern for focusing a plane wave, 𝑟 2 𝑛 𝑛 = 𝑛 𝜆 𝑓 + 2 𝜆 2 4 , ( 2 ) it is required that f is much greater than 𝑛 𝜆 / 4 for the ignorable spherical aberrations [12]. According to the Rayleigh’s quarter wavelength rule [13], the spherical aberration can be ignored, when the following condition is satisfied [8]: 𝑁 2 2 < 𝑓 𝜆 . ( 3 ) Here, N is total number of zone. In order to satisfy the Rayleigh’s 𝜆 /4 rule, we chose N = 1265, that is, 𝑟 𝑁 = 88.56  𝜇 m (diameter of 177.12  𝜇 m). Then, the outermost zone width becomes 35.00 nm. The 𝜆 /4 rule is satisfied in the wavelength region shorter than 2.8 nm. Thus, the spherical aberration can be ignored in the hard X-ray region.

The theoretical spatial resolution (Rayleigh resolution limit) for the first-order diffraction of FZP is known to be 1 . 2 2 𝑑 𝑁 , where 𝑑 𝑁 is the width of outermost zone [12]. Therefore, if the zone structure is ideal, spatial resolution of the FZP could be 43 nm. It should be noted that the knife-edge scan method, which is generally used for measuring the focused beam shape, gives a line spread function of beam profile. The line-spread-function, L S F ( 𝑥 ) , of a diffraction-limited focal spot for circular aperture lens is given by one-dimensional integral of point spread function as L S F ( 𝑥 ) = 2 𝐽 1 𝑥 2 + 𝑦 2 2 𝑥 2 + 𝑦 2 𝑑 𝑦 , ( 4 ) where 𝐽 1 ( 𝑥 ) is a Bessel function of the first kind. The Rayleigh’s criterion for spatial resolution is defined by the first minimum of the Bessel function as 𝐽 1 ( 𝑥 1 ) = 0 , and the solution of 𝑥 1 = 1 . 2 2 0 𝜋 is usually used as a definition of diffraction-limited resolution [13], while the full width at half maximum (FWHM) of point-spread-function, { 2 𝐽 1 ( 𝑥 ) / 𝑥 } 2 , is 1.0290 𝜋 . It is shown by the numerical calculation of the above formula that the FWHM of line-spread-function is given by 𝑥 = 1 . 0 0 1 6 𝜋 . Therefore, the FWHM of line-spread-function is nearly equal to the outermost zone width of ideal FZP, and the theoretical FWHM of knife-edge scan profile for the diffraction-limited micro focus beam is 1 . 0 0 1 6 𝑑 𝑁 .

The calculated diffraction efficiency for the 0.175  𝜇 m-thick tantalum FZP is only 0.8% for 8 keV X-rays. This low efficiency may sometimes be a problem for nanobeam applications or for imaging microscopy because of low signal intensity. However, the low diffraction efficiency can be compensated by high brilliance X-ray sources. Some scanning microscopy experiments can be carried out with this zone plate, and the application to the imaging microscopy would also be capable at the expense of relatively long exposure time and high radiation dose to specimen. This low diffraction efficiency is not a problem for characterization of FZP.

The thickness of FZP (0.175  𝜇 m) is, however, not optimized for 8 keV X-rays. The optimum thickness of Ta zone plate for 8 keV is 1.9  𝜇 m. Therefore, higher aspect ratio of zone structure is required for improvement of diffraction efficiency. The aspect ratio of tantalum one-to-one zone structure is presently limited by technological reason, mainly for reactive dry-etching process. The present limit of aspect ratio is about 10. The aspect ratio doubling by stacking two zone plates [7] would be a promising method for higher diffraction efficiency. Zone material of high electron density is adequate for X-ray zone plates, because the optimized thickness is in proportion to the reciprocal of electron density. Gold is generally chosen as the zone material for its high density and fabrication technology, for example, electroplating of gold. The density of tantalum is less than that of gold by about 14%. Therefore, the zone plate made of  Ta usually gives lower efficiency than that made of Au in the case of same zone thickness. However, we consider that the difference of diffraction efficiency is not significant, and the tantalum is also one of the best materials for hard X-ray zone plates. In some cases, for example, near the absorption edge of gold, tantalum zone plate with the same thickness can give slightly higher efficiency than gold zone plate.

3. Performance Test of FZP

Performance test of the FZP as a nanofocusing optics has been carried out at the beamline 20XU [14] of SPring-8. A schematic view of the experimental setup is shown in Figure 3. An X-ray energy of 8 keV was chosen in the experiment. The energy resolution of monochromator is about 1.1 eV for 8 keV X-rays. To avoid the chromatic aberration of FZP optics, the relative spectral bandwidth of incident X-ray beam, Δ 𝜆 / 𝜆 , must be less than or equal to the inverse of total number of zone [12], as Δ 𝜆 𝜆 1 𝑁 . ( 5 ) Comparing with the zone number of the FZP, 𝑁 = 1265, the monochromaticity of Δ E/E ~ 1/7000 is sufficient for testing the performance of FZP.

The monochromatic beam is transported through the 200-m-long vacuum duct and is extracted in to the air at an end experimental station located at approximately 245 m from the light source. A precise cross slit located 49 m from the source is used as a pseudopoint source for determining the spatial coherence at the end station. The cross slit with an opening of 50  𝜇 m × 50  𝜇 m was used in the present experiment. Therefore, the geometrical spot size for an FZP with a focal length of 40 mm becomes 10 nm that is much less than the diffraction-limited resolution of the FZP. There are no optical elements between the cross-slit virtual source and the experimental station, except for a beamline X-ray window made of 125  𝜇 m-thick polyimide.

The focus spot profile is evaluated by knife-edge scan method. However, it is difficult to fabricate precise knife-edge for characterization of nano-focusing beam, because of high transmissivity of hard X-ray beam. So, the conventional knife-edge test cannot be used for the present experiment. We employed dark-field method for the knife-edge test [1]. A pattern edge of tantalum resolution test chart is used as a knife-edge. The thickness of tantalum test pattern is 0.5  𝜇 m, which corresponds to an optical path difference of about 𝜆 /8 at an X-ray wavelength of 0.155 nm. The intensity of the diffracted wave takes its maximum at an optical thickness of 𝜆 /2 for phase edge. Therefore, from point of view of signal intensity, the optimum thickness of knife-edge for dark-field method is 𝜆 /2. However, the thickness of 𝜆 /8 is still sufficient for the beam profile measurement. The thicker knife-edge may give higher dark-field signal, but the thickness must be much less than the depth of focus that is 7.9  𝜇 m in the present experimental condition. The test pattern edge was set parallel to the optical axis in order to avoid broadening of edge-scan profiles. The misalignment of substrate of test pattern was within an angle of about 0.3 degree.

The knife-edge scanning was carried out using a conventional mechanical stage driven by an open-loop-controlled step motor. The accuracy of the translation stage is measured to be better than 10 nm [15]. The X-ray beam intensity was measured with ionization chamber detectors with 14 cm electrode length. One is an air-flowed ionization chamber placed in front of the FZP, and the other is an argon-flowed ionization chamber behind the knife-edge. A beam stop is placed in front of the argon-flowed ionization chamber for the purpose of dark-field edge scan.

A typical example of measured beam profiles is shown in Figure 4. The FWHM of the focused beam profile shown in Figure 4 is 34.4 nm. The accuracy of beam size measurement was estimated from repeatability of measured beam size. The average of thirteen measurements around the focus is 34.9 nm, and the error of beam size measurement is 2.7 nm in standard deviation. Therefore, it is concluded that the focused beam size is 34.9 nm ± 2.7 nm in FWHM. As shown in the figure, the measured beam profile agrees well with the theoretical one. It is concluded that nanometer-resolution imaging is capable in the hard X-ray region by using the FZP as a beam-focusing device or as an objective lens for imaging microscopy. The measured diffraction efficiency of the FZP is 0.7% for 8 keV X-rays. This value agrees well with the theoretical efficiency (0.8%) of 0.175  𝜇 m-thick Ta zone plate. Modulation transfer function (MTF) obtained from the measured beam profile in Figure 4 is shown in Figure 5. The Rayleigh’s resolution limit (1.22 𝑑 𝑁 ) for 35 nm outermost zone width that corresponds to a spatial frequency of 23.4  𝜇 m-1 is indicated by an arrow in the figure. The measured MTF at 23.4 cycle/ 𝜇 m is about 10%.

4. Scanning Microscopy Experiments

Experimental setup for scanning microscopy experiment is shown in Figure 6. Bright field imaging and dark-field imaging [16] can be exchanged by rotating the beam stop in front of the detector. The beam stop is made of a lead plate with a thickness of 0.5 mm. Scanning microscopy images of a test object are shown in Figures 7(a) and 7(b). The specimen is resolution test patterns made of 0.5  𝜇 m-thick tantalum that were also fabricated at NTT-AT Nanofabrication by the same process as that of the FZP. The design feature of line and space patterns is shown in Figure 7(c). The fine patterns up to 0.2  𝜇 m were clearly seen in the measured images. The beam flux at the sample position was about 4 × 106 photons/s, and signal intensity in dark field mode was 2 × 105 photons/s.

Although the design of test pattern has the 0.1  𝜇 m-wide line and space patterns, the 0.1  𝜇 m wide patterns are not observed in the measure images, and a weak line is recognized at the right edge of pattern in the dark-field image where the 0.1  𝜇 m-wide line should be seen. The right edge of the 0.2  𝜇 m-line pattern is not observed in the dark-field image. In addition, in the bright-field image, there is a faint contrast pattern at the right side of the 0.2  𝜇 m-line pattern. From these results, we consider that the 0.1  𝜇 m-wide pattern is provably collapsed or slanted and contacts to the neighboring 0.2  𝜇 m-wide line pattern. The 0.1  𝜇 m-wide line and space patterns are unfortunately imperfect in the measured portion of large test patterns.

Acknowledgment

The experiments were performed at the BL20XU in SPring-8 under the approval of the Japan Synchrotron Radiation Research Institute (JASRI) (Proposal no. 2006A1562).