Research Article

Speckle Reduction for Ultrasonic Imaging Using Frequency Compounding and Despeckling Filters along with Coded Excitation and Pulse Compression

Table 2

CNR, HO, , and cSNR for the 15 mm ATS phantom target.

Technique1 CNR HO cSNR

CP 0.56 32.49 16.06ā€”
REC 0.62 32.49 17.61 8.23
REC-FC (third-width) 1.73 14.07 43.17 12.82
eREC-FC 1.60 17.22 37.64 8.72
eREC-FC and median filtering 2.29 12.12 45.43 8.08
eREC-FC and Lee filtering 2.57 12.09 44.89 8.14
eREC-FC and HMA filtering 2.51 13.41 43.98 8.12
eREC-FC and geometric filtering 1.83 15.17 37.67 10.07
eREC-FC and SRAD filtering 1.86 10.65 40.49 18.32

CP: conventional pulsing; REC: resolution enhancement compression; FC: frequency compounding; eREC-FC: enhanced REC-FC; HMA: homogeneous mask area filtering; SRAD: speckle reducing anisotropic diffusion.