Research Article
Speckle Reduction for Ultrasonic Imaging Using Frequency Compounding and Despeckling Filters along with Coded Excitation and Pulse Compression
Table 2
CNR, HO,
, and cSNR for the 15 mm ATS phantom target.
| Technique1 | CNR | HO | | cSNR |
| CP | 0.56 | 32.49 | 16.06 | ā | REC | 0.62 | 32.49 | 17.61 | 8.23 | REC-FC (third-width) | 1.73 | 14.07 | 43.17 | 12.82 | eREC-FC | 1.60 | 17.22 | 37.64 | 8.72 | eREC-FC and median filtering | 2.29 | 12.12 | 45.43 | 8.08 | eREC-FC and Lee filtering | 2.57 | 12.09 | 44.89 | 8.14 | eREC-FC and HMA filtering | 2.51 | 13.41 | 43.98 | 8.12 | eREC-FC and geometric filtering | 1.83 | 15.17 | 37.67 | 10.07 | eREC-FC and SRAD filtering | 1.86 | 10.65 | 40.49 | 18.32 |
|
|
CP: conventional pulsing; REC: resolution enhancement compression; FC: frequency compounding; eREC-FC: enhanced REC-FC; HMA: homogeneous mask area filtering; SRAD: speckle reducing anisotropic diffusion.
|