Mechanical Testing of a Novel Fastening Device to Improve Scoliosis Bracing Biomechanics for Treating Adolescent Idiopathic Scoliosis

Table 1

Corrective forces and moments applied by brace to the SAM.

Brace configurationUpper load cell readingsLower load cell readings
BracePosterior strapsDynamic loadFx (N)Fy (N)Fz (N)Mt (Nm)Fx (N)Fy (N)Fz (N)Mt (Nm)

CBVelcroN/A−15.10.4−175.4−0.711.220.1−128.70.1
CBCTUN/A−16.01.0−191.6−0.514.818.3−141.80.1
MBVelcro0 N−12.1−8.7−81.2−1.36.17.8−39.40.1
MBVelcro35.6 N−10.8−7.4−93.9−0.75.111.4−60.20.1
MBVelcro71.2 N−13.54.5−180.3−1.19.025.1−129.20.1
MBCTU0 N−14.63.5−169.0−0.611.623.2−126.10.1
MBCTU35.6 N−15.210.0−231.6−0.611.933.8−177.90.1
MBCTU71.2 N−15.711.2−220.4−0.413.731.9−169.70.1