Research Article

Surface and Buildup Region Dose Measurements with Markus Parallel-Plate Ionization Chamber, GafChromic EBT3 Film, and MOSFET Detector for High-Energy Photon Beams

Table 1

(a) Percentage depth doses (PDDs) obtained with a Markus chamber in a water equivalent RW3 phantom using 6 MV photon beams for different field sizes at SSD = 100 cm. (b) Percentage depth doses (PDDs) obtained with a Markus chamber in a water equivalent RW3 phantom using 15 MV photon beams for different field sizes at SSD = 100 cm.
(a)

Phantom depth (mm)WET (mm) PDD (%)
5 × 5 cm210 × 10 cm220 × 20 cm2

00.02310.8 ± 0.816.6 ± 0.928.1 ± 0.8
0.0670.0714.6 ± 0.920.3 ± 0.931.4 ± 0.8
11.06838.7 ± 0.843.4 ± 0.852.5 ± 0.7
22.11357.1 ± 0.761.0 ± 0.668.2 ± 0.6
55.24883.6 ± 0.685.9 ± 0.490.0 ± 0.4
1010.47397.8 ± 0.298.4 ± 0.299.2 ± 0.3
1515.698100.0 ± 0.1100.0 ± 0.1100.0 ± 0.2

(b)

Phantom depth (mm)WET (mm) PDD (%)
5 × 5 cm210 × 10 cm220 × 20 cm2

00.0236.9 ± 0.913.7 ± 0.826.7 ± 0.8
0.0670.078.0 ± 0.914.9 ± 0.827.9 ± 0.8
11.06822.7 ± 0.829.6 ± 0.742.0 ± 0.8
22.11334.8 ± 0.741.5 ± 0.653.5 ± 0.9
55.24860.3 ± 0.865.8 ± 0.775.6 ± 0.8
1010.47381.3 ± 0.685.4 ± 0.592.1 ± 0.6
1515.69892.3 ± 0.494.8 ± 0.398.6 ± 0.4
2020.92397.1 ± 0.398.7 ± 0.3101.0 ± 0.9
2526.14899.6 ± 0.3100.1 ± 0.3100.9 ± 0.7
3031.373100.0 ± 0.2100.0 ± 0.2100.0 ± 0.3