Surface and Buildup Region Dose Measurements with Markus Parallel-Plate Ionization Chamber, GafChromic EBT3 Film, and MOSFET Detector for High-Energy Photon Beams

Table 3

(a) Percentage depth doses (PDDs) obtained with a MOSFET dosimeter in a water equivalent RW3 phantom using 6 MV photon beams for different field sizes at SSD = 100 cm. (b) Percentage depth doses (PDDs) obtained with a MOSFET dosimeter in a water equivalent RW3 phantom using 15 MV photon beams for different field sizes at SSD = 100 cm.
(a)

Phantom depth (mm)WET (mm) PDD (%)
5 × 5 cm210 × 10 cm220 × 20 cm2

N/A0.0717.4 ± 1.325.5 ± 1.334.5 ± 1.8
01.80040.1 ± 1.446.4 ± 1.657.2 ± 1.6
12.84556.8 ± 1.060.6 ± 1.569.3 ± 1.5
23.89066.5 ± 1.270.2 ± 1.677.2 ± 1.6
57.04587.8 ± 1.189.0 ± 1.497.5 ± 1.7
1012.250106.3 ± 0.8105.2 ± 1.2105.5 ± 1.9
1517.475100.0 ± 0.9100.0 ± 1.1100.0 ± 1.8

(b)

Phantom depth (mm)WET (mm) PDD (%)
5 × 5 cm210 × 10 cm220 × 20 cm2

N/A0.079.0 ± 1.616.4 ± 1.928.9 ± 1.8
01.80024.6 ± 1.733.0 ± 1.940.0 ± 1.7
12.84530.0 ± 1.640.0 ± 1.852.4 ± 1.6
23.89042.3 ± 1.550.3 ± 1.659.1 ± 1.6
57.04562.1 ± 1.670.0 ± 1.774.8 ± 1.4
1012.25079.4 ± 1.788.7 ± 1.690.8 ± 1.5
1517.47589.9 ± 1.397.9 ± 1.595.2 ± 1.6
2022.70096.7 ± 1.299.0 ± 1.699.2 ± 1.8
2527.92598.3 ± 1.399.3 ± 1.799.8 ± 1.8
3033.150100.0 ± 1.0100.0 ± 1.2100.0 ± 1.1