Research Article
Forecasting of Corrosion Properties of Steel Wires for Production of Guide Wires for Cardiological Treatment
Table 3
Test results of pitting corrosion resistance.
| Wire diameter , mm | Strain in drawing process | Breakdown potential , mV | Polarisation resistance , kcm2 | Corrosion current density , A/cm2 |
| Electrochemically polished wires | 5.6 | 0 | +612 | 567 | 0.04 | 3.0 | 1.27 | +510 | 459 | 0.10 | 1.5 | 2.65 | +380 | 229 | 0.11 |
| Electrochemically polished and chemically passivated wires | 5.6 | 0 | +858 | 2220 | 0.012 | 3.0 | 1.27 | +778 | 1230 | 0.021 | 1.5 | 2.65 | +695 | 1070 | 0.024 |
|
|