Potentiometric Response of Ag/AgCl Chloride Sensors in Model Alkaline Medium

Table 2

Model solutions and criteria for sensors’ performance.

Test seriesUsed mediumpHChloride concentration (mM)Purpose of OCP measurement in this environment (criteria for sensors’ performance)

1Simulated pore solution (SPS)13.60Stability: OCP variation in chloride-free alkaline solution (Figure 2)
2Cement extract solution (CE)12.820, 260Reproducibility: dependent on different properties of the AgCl layer (Figure 4)
3Simulated pore solution (SPS)7, 9, 12.8, 13.61, 2, 4, 16, 62, 250Reliability: interference of hydroxide ions and chloride detection limits (Figure 6)
4Cement extract solution (CE)12.812, 14, 16, 20, 27, 44, 77, 142, 260, 513, 1008Reversibility: response to continuous addition of chloride (Figures 7 and 8)
5Simulated pore solution (SPS)13.60, 125, 1000