Detection of Nitroaromatic and Peroxide Explosives in Air Using Infrared Spectroscopy: QCL and FTIR

Table 1

Validation parameters for the various models constructed.

ParameterModulated FTIR (model 1)Unmodulated FTIR (model 2)QCL scan (model 3)
TATPDNTTATPDNTTATPDNT

Wilks’ lambda0.120.030.090.110.050.06
Canonical correlation0.940.980.960.940.980.97
PCCC100.00%100.00%100.00%100.00%100.00%100.00%
PCCCC100.00%99.94%100.00%99.97%100.00%100.00%
Sensibility100.00%100.00%100.00%99.50%100.00%100.00%
Specificity100.00%99.75%100.00%100.00%100.00%100.00%
False alarm0.00%0.00%0.00%0.50%0.00%0.00%
Missed detection0.00%0.25%0.00%0.00%0.00%0.00%
Loadings77101055
Samples ( )30793079300300106106
value<0.0001<0.0001<0.0001<0.0001<0.0001<0.0001