Research Article

Experiments on the Porch Swing Bearing of Michelson Interferometer for Low Resolution FTIR

Figure 8

The measured and calculated values of the pitch tilt as a function of the length difference between the arms. The pitch was caused by the length difference of the arms. The error bars represent the maximum measurement errors estimated from the readings of the scales on the measurement equipment.