Clinical Study

Randomised Comparison of the AMBU AuraOnce Laryngeal Mask and the LMA Unique Laryngeal Mask Airway in Spontaneously Breathing Adults

Table 2

Results.

LMAU ( )AMBU ( ) -value

Insertion success
 First time31 (82)35 (85)0.738
 Overall38 (100)40 (98)0.519
Insertion time (s) 3216 (11–95) 2710 (13–51) 0.111
OLP (cm H2O) 157 (4–30) 206 (5–30) 0.001
Laryngeal Alignment
 I21 (55)21 (54)
 II9 (24)9 (23)
 III7 (18)7 (18)
 IV1 (3)2 (5)0.613
Blood Staining 3 (8) 4 (10) 0.543
Sore Throat PACU 4 (11) 3 (8) 0.457

All data are presented as mean + standard deviation (range) or numbers (%). LMAU: LMA Unique laryngeal mask, AMBU: AMBU AuraOnce laryngeal mask, OLP: oropharyngeal leakage pressure, laryngeal alignment (I: visualise entire vocal cords, II: posterior component of cords visible, III: epiglottis but no cords visible, IV: no vocal cords or epiglottis visible).