Evaluation of Satisfaction and Axial Rigidity with Titan XL Cylinders
Table 5
Rigidity measurements.
| | Patient | Physician |
| Number of pumps until perceiving full inflation | | | N | 16 | 16 | Mean ± SD | | | Median (Min, Max) | 28.5 (9, 58) | 44.5 (33, 62) | 95% CI | 22.6, 38.2 | 40.0, 48.9 | Rigidometer reading (g) | | | N | 16 | 16 | Mean ± SD | | | Median (Min, Max) | 1600 (800, 2000) | 1800 (1400, 2000) | 95% CI | 1317.5, 1682.5 | 1674.3, 1900.7 | Did the penis buckle? | | | No (n/N (%)) | 5/16 (31%) | 16/16 (100%) | Yes (n/N (%)) | 11/16 (69%) | 0/16 (0%) |
|
|