Internally Controlled, Generic Real-Time PCR for Quantification and Multiplex Real-Time PCR with Serotype-Specific Probes for Serotyping of Dengue Virus Infections
Figure 1
The dynamic range of generic of PCR. Mean Cq-values of twelve replicates per each 10-fold dilutions of DENV in vitro RNA extracted with a fixed amount of IC copies in extraction in the back ground of DENV negative serum and tested in generic real-time PCR resulting in a dynamic range of –6.25 × 107 copies/mL of DENV in vitro RNA with a regression coefficient of 0.9967.