Volume 2013 (2013), Article ID 216034, 12 pages
http://dx.doi.org/10.1155/2013/216034
Research Article

Total Body Irradiation with Step Translation and Dynamic Field Matching

1Department of Radiation Oncology, Taichung Veterans General Hospital, 1650 Taiwan Boulevard Sect. 4, Taichung 40705, Taiwan
2Department of Biomedical Engineering and Environmental Sciences, National Tsing Hua University, Taiwan
3Department of Biomedical Imaging and Radiological Science, China Medical University, Taichung 40402, Taiwan

Received 14 April 2013; Accepted 31 May 2013

Academic Editor: Chung-Chi Lee

Copyright © 2013 Ho-Hsing Chen et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

The purpose of this study is to develop a total body irradiation technique that does not require additional devices or sophisticated processes to overcome the space limitation of a small treatment room. The technique aims to deliver a uniform dose to the entire body while keeping the lung dose within the tolerance level. The technique treats the patient lying on the floor anteriorly and posteriorly. For each AP/PA treatment, two complementary fields with dynamic field edges are matched over an overlapped region defined by the marks on the body surface. A compensator, a spoiler, and lung shielding blocks were used during the treatment. Moreover, electron beams were used to further boost the chest wall around the lungs. The technique was validated in a RANDO phantom using GAFCHROMIC films. Dose ratios at different body sites along the midline ranged from 0.945 to 1.076. The dose variation in the AP direction ranged from 96.0% to 104.6%. The dose distribution in the overlapped region ranged from 98.5% to 102.8%. Lateral dose profiles at abdomen and head revealed 109.8% and 111.7% high doses, respectively, at the body edges. The results confirmed that the technique is capable of delivering a uniform dose distribution to the midline of the body in a small treatment room while keeping the lung dose within the tolerance level.