Injection time (s) | Concentration (%) | Injection rate (mL/s) | Effects and analysis of perfusion imaging | Methods to improve image quality |
|
1 | 19–76 | 0.3–0.5 | Low concentration, short duration, and no pseudo-color map | Extend the injection time |
1 | 76 | 0.5 | High concentration, slow injection rate, and no pseudo-color map | Unfeasible method |
2~3 | 76 | 0.3 | Short duration and no pseudo-color map | Extend the injection time |
2~3 | 19–57 | 0.3–0.5 | Low concentration, short duration, and no pseudo-color map | Extend the injection time |
2~3 | 76 | 0.5 | High concentration, slow injection, and no pseudo-color map | Unfeasible method |
4~5 | 76 | 0.3 | Short duration and no pseudo-color map | Extend the injection time |
4~5 | 19–38 | 0.3–0.5 | Low concentration, short duration, and no pseudo-color map | Extend the injection time |
4~5 | 57 | 0.5 | Poor quality of pseudo-color map | Extend the injection time |
4~5 | 76 | 0.5 | High concentration, slow injection, and no pseudo-color map | Unfeasible method |
6 | 76 | 0.3 | Poor quality of pseudo-color map | Decrease concentration |
6 | 19–38 | 0.3–0.5 | Low concentration and no pseudo-color map | Increase concentration |
6 | 57 | 0.5 | Good pseudo-color map | Feasible method |
6 | 76 | 0.5 | High concentration, slow injection, and no pseudo-color map | Unfeasible method |
|