Use of Tekscan K-Scan Sensors for Retropatellar Pressure Measurement Avoiding Errors during Implantation and the Effects of Shear Forces on the Measurement Precision

Table 2

Percentage decrease of the measured retropatellar parameters in 12 consecutive measurement cycles without changing the experimental setup.

Torque 30 Nm
Cyclus 1Cyclus 2Cyclus 3Cyclus 7Cyclus 8Cyclus 9

Force0% −1.6% ± 1.3−3.3% ± 1.4−9.5% ± 4.7−10.6% ± 4.1−12.3% ± 3.9
Area0%−0.7% ± 1.0−1.5% ± 0.9−5.9% ± 4.2−6.2% ± 3.4−7.6% ± 2.9
Pressure0%−1.0% ± 0.7−1.8% ± 1.0−3.8% ± 2.1−4.6% ± 2.0−5.0% ± 2.1
Max. force0%−1.2% ± 1.0−2.2% ± 1.3−3.8% ± 2.2−4.5% ± 2.4−5.1% ± 2.4
Max. pressure0%−1.1% ± 0.9−2.1% ± 1.2−3.8% ± 2.2−4.4% ± 2.4−5.0% ± 2.4

Torque 40 Nm
Cyclus 4Cyclus 5Cyclus 6Cyclus 10Cyclus 11Cyclus 12

Force0%−0.8% ± 1.0−2.6% ± 1.9−8.1% ± 3.1−8.7% ± 1.8−10.1% ± 3.1
Area0%−0.7% ± 0.9−1.8% ± 2.1−4.2% ± 2.4−5.2% ± 2.3−6.2% ± 1.7
Pressure0%−0.2% ± 0.8−1.0% ± 0.8−3.7% ± 2.2−3.8% ± 1.8−4.5% ± 2.6
Max. force0%−0.3% ± 1.7−1.2% ± 2.1−2.8% ± 1.4−3.5% ± 1.9−4.2% ± 2.9
Max. pressure0%−0.3% ± 1.6−1.2% ± 2.1−2.8% ± 1.4−3.4% ± 1.9−4.2% ± 2.9