Removal of Cardiopulmonary Resuscitation Artifacts with an Enhanced Adaptive Filtering Method: An Experimental Trial
Table 1
Estimated signal-to-noise ratio (SNR) for pulseless electrical activity (PEA), ventricular fibrillation (VF), and asystole (ASY) before and after filtering.
Unfiltered
Adaptive filter
High-pass filter
Medians (dB) (25/75 percentiles)
VF
−9.3 (−14.9/−3.6)△△
0.2 (−5.1/4.5)**
0.1 (−4.2/0.9)**
PEA
−6.2 (−9.0/−1.12)△△
0.1 (−3.6/3.4)**
−2.0 (−7.4/−0.6)**
ASY
−21.2 (−24.2/−18.5)△△
−12.7 (−15.0/−4.4)**
−7.1 (−10.7/−6.3)**
Range (dB) (min./max.)
VF
−26.1/9.6
−18.2/20.0
−19.7/20.4
PEA
−16.0/9.9
−7.6/19.9
−14.0/14.7
ASY
−31.6/−10.0
−20.6/2.4
−18.4/1.7
Compared with unfiltered signal, ; comparison among rhythm types, .