Comparison of Diagnostic Accuracy between Octopus 900 and Goldmann Kinetic Visual Fields
Table 2
Mean test durations and mean differences.
| Perimeter option | Mean test duration (minutes) | SD |
| Goldmann | 2.73 | 0.71 | Octopus 3°/sec | 4.46 | 1.42 | Octopus 5°/sec | 2.33 | 0.55 | Octopus 10°/sec | 2.30 | 0.62 |
| Perimeter comparison | Mean differences (minutes) | Significance (ANOVA: Bonferroni correction) |
| Goldmann-Octopus 3°/sec | 1.76 | | Goldmann-Octopus 5°/sec | −0.36 | not significant | Goldmann-Octopus 10°/sec | −0.46 | | Octopus 3–5°/sec | −2.04 | | Octopus 3–10°/sec | −2.33 | | Octopus 5–10°/sec | −0.09 | not significant |
|
|