The Role of Imaging in Radiation Therapy Planning: Past, Present, and Future

Table 2

Sources of MR distortion, correction, and method for assessment.

Level of distortionDistortion sourceDistortion correction Assessment of residual distortion

System Gradient nonlinearityGradient coil design (HW) 
Model-based corrections (SW)
Geometric fidelity phantoms and assessment tools
Static field (B0) 
inhomogeneity
Magnet design and static-field shims (HW)B0-field map in phantoms

Patient Static-field (B0) 
inhomogeneity
Gradient-shim coils (HW) 
Patient-specific B0 field correction (SW)
B0-field map in the patient
Transmit-field (B1) 
inhomogeneity
Multiple RF transmit channels (HW) 
Patient-specific B1 field correction (SW)
B1-field map in the patient

Sequence Gradient switching 
(eddy currents)
Gradient preemphasis (HW/SW) 
Protocols with low gradient slew rates (SW)
Visual inspection