Clinical Study

Four-Dimensional Computed Tomography Based Respiratory-Gated Radiotherapy with Respiratory Guidance System: Analysis of Respiratory Signals and Dosimetric Comparison

Table 2

The root mean square error (RMSE) of maxima, minima, and period in the three breathing conditions.
(a) The RMSE of maxima (full inspiration)

Free Audio-guidedComplex system-guidedP value

Maxima (cm)0.154 ± 0.0960.119 ± 0.0570.061 ± 0.0240.031

Free breathing versus audio-guided breathing (P = 0.216).
Free breathing versus complex system-guided breathing (P = 0.004).
Audio-guided breathing versus complex system-guided breathing (P = 0.001).
(b) The RMSE of minima (full expiration)

FreeAudio-guidedComplex system-guidedP value

Minima (cm)0.220 ± 0.0850.174 ± 0.0950.099 ± 0.0530.007

Free breathing versus audio-guided breathing (P = 0.020).
Free breathing versus complex system-guided breathing (P = 0.000).
Audio-guided breathing versus complex system-guided breathing (P = 0.020).
(c) The RMSE of period

FreeAudio-guidedComplex system-guidedP value

Period (sec)0.951 ± 0.1300.741 ± 0.2130.464 ± 0.2150.046

Free breathing versus audio-guided breathing (P = 0.370).
Free breathing versus complex system-guided breathing (P = 0.009).
Audio-guided breathing versus complex system-guided breathing (P = 0.040).