Comparison of Anterior Segment Measurements Obtained by Three Different Devices in Healthy Eyes

Figure 5

Bland-Altman analysis showing the distribution of the WTW differences (WTWiol-WTWgal) on the -axis and the average of the instrument readings (WTWiol + WTWgal)/2 on the -axis. Mean : −0.080 ± 0.274 mm, with 95% limits of agreement between −0.628 and +0.467.