Comparison of Anterior Segment Measurements Obtained by Three Different Devices in Healthy Eyes

Table 1

Mean values and variance of the flat SimK, steep SimK, corneal astigmatism, ACD, and WTW measurements by the IOL Master, Galilei, and WAM 5500 methods and the intraclass coefficients for three repeated measurements.

Mean (±)ICCLower 95% CIUpper 95% CICOV (%)

IOL MasterFlat SimK (D)43.531.490.9990.9990.9993.43
Steep SimK (D)44.281.520.9990.9980.9993.44
Corneal astigmatism (D)0.760.420.9710.9590.98055.35
ACD (mm)3.600.300.9740.9650.9828.23
WTW (mm)12.120.420.9670.9530.9783.47

GalileiFlat SimK (D)43.371.480.9970.9950.9983.42
Steep SimK (D)44.111.510.9980.9970.9993.43
Corneal astigmatism (D)0.740.400.8550.7930.90153.54
ACD (mm)3.870.420.9370.9080.9583.42
WTW (mm)12.200.460.9370.9080.9583.79

WAM 5500Flat SimK (D)43.521.490.9920.9880.9953.44
Steep SimK (D)44.341.590.9880.9810.9923.58
Corneal astigmatism (D)0.820.480.7900.6790.86258.78

: within-subjects variance; ICC: intraclass coefficients; CI: confidence interval; COV: coefficient of variation.