Research Article

Comparison of Anterior Segment Measurements Obtained by Three Different Devices in Healthy Eyes

Table 2

The mean differences, standard deviation, and Wilcoxon test (values of P < 0.05 were considered to be indicative of significant differences) of the flat SimK, steep SimK, corneal astigmatism, ACD, and WTW measurements by the IOL Master, Galilei, and WAM 5500.

Mean differenceSD (±)P

Flat SimK (D)IOL Master-Galilei0.1550.156<0.001
IOL Master-WAM 55000.0070.1870.302
Galilei-WAM 5500−0.1480.195<0.001

Steep SimK (D)IOL Master-Galilei0.1710.168<0.001
IOL Master-WAM 5500−0.0550.2270.172
Galilei-WAM 5500−0.2260.236<0.001

Corneal astigmatism (D)IOL Master-Galilei0.0160.2490.662
IOL Master-WAM 5500−0.0610.2650.055
Galilei-WAM 5500−0.0780.2920.022

ACD (mm)IOL Master-Galilei−0.2690.158<0.001

WTW (mm)IOL Master-Galilei−0.0800.2740.009