Research Article
Comparison of Anterior Segment Measurements Obtained by Three Different Devices in Healthy Eyes
Table 3
Spearman correlation coefficients between devices.
| ā | ā | Ro Spearman | P |
| Flat SimK (D) | IOL Master-Galilei | 0.992 | <0.001 | IOL Master-WAM 5500 | 0.990 | <0.001 | Galilei-WAM 5500 | 0.986 | <0.001 |
| Steep SimK (D) | IOL Master-Galilei | 0.991 | <0.001 | IOL Master-WAM 5500 | 0.986 | <0.001 | Galilei-WAM 5500 | 0.984 | <0.001 |
| Corneal astigmatism (D) | IOL Master-Galilei | 0.777 | <0.001 | IOL Master-WAM 5500 | 0.792 | <0.001 | Galilei-WAM 5500 | 0.754 | <0.001 |
| ACD (mm) | IOL Master-Galilei | 0.809 | <0.001 |
| WTW (mm) | IOL Master-Galilei | 0.791 | <0.001 |
|
|