Clinical Study

Temporal Parameters Estimation for Wheelchair Propulsion Using Wearable Sensors

Table 6

Stroke number and push frequency intraclass correlation coefficient (ICC 3, 1).

ā€‰ā€‰ICC95% CI value

Stroke numberARM0.994 0.988~0.997<0.001
WRIST0.9900.980~0.995<0.001
SEAT0.9840.972~0.991<0.001

Push frequencyARM0.9160.843~0.953<0.001
WRIST0.8890.802~0.936<0.001
SEAT0.6900.071~0.868<0.001

CI: confidence interval.