Clinical Study

Repeatability and Reproducibility of Retinal Nerve Fiber Layer Parameters Measured by Scanning Laser Polarimetry with Enhanced Corneal Compensation in Normal and Glaucomatous Eyes

Table 4

Intravisit repeatability and intervisit reproducibility of SLP-ECC parameters in the glaucoma group ().

GDx parametersIntravisitIntervisit
ICCICC 95% CICOV
(%)
TRT
SD
ICCICC 95% CICOV
(%)
TRT
SD
Upper limitLower limitUpper limitLower limit

NFI0.9820.9720.989<0.00119.3810.610.9750.9580.985<0.00115.8212.84
TSNIT average0.9770.9610.986<0.0014.403.530.9770.9610.986<0.0014.713.93
Superior average0.9200.8780.950<0.0017.426.650.9170.8630.952<0.0017.886.86
Inferior average0.9600.9390.975<0.0015.685.480.9780.9640.987<0.0015.365.13
TSNIT SD 0.9510.9240.969<0.0019.113.230.9700.9510.983<0.0018.302.83

ICC: intraclass correlation coefficient; CI: confidence interval; COV: coefficient of variation; TRT SD: test-retest variability; NFI: nerve fiber indicator; TSNIT: temporal-superior-nasal-inferior-temporal; SD: standard deviation.