Testing of different qRT-PCR kits (one-step)
Number of copies/sampleNumber of copies/rxnQuantitect-1 stepRotor Gene-1 stepVerso-1 stepSSIII Platinium-1 stepqScript-1 step
Average CtCV (%)Average CtCV (%)Average CtCV (%)Average CtCV (%)Average CtCV (%)

1062.0 10616.66 ± 0.08 0.45 17.15 ± 0.05 0.29 16.99 ± 0.06 0.38 14.52 ± 0.19 0.32 13.47 ± 0.03 0.19
17.79 ± 0.241.3215.88 ± 0.221.3618.34 ± 0.321.7216.22 ± 0.040.2218.48 ± 0.040.20
1052.0 10520.06 ± 0.10 0.52 20.62 ± 0.04 0.17 20.40 ± 0.11 0.52 17.90 ± 0.21 1.18 16.55 ± 0.03 0.19
21.13 ± 0.452.1219.42 ± 0.201.0421.60 ± 0.150.7119.64 ± 0.412.1122.38 ± 0.150.68
1042.0 10423.58 ± 0.14 0.58 23.91 ± 0.08 0.33 23.82 ± 0.20 0.83 21.43 ± 0.12 0.57 19.86 ± 0.04 0.18
24.60 ± 0.200.8322.87 ± 0.160.6925.17 ± 0.170.6923.08 ± 0.271.1626.11 ± 0.100.38
1032.0 10326.70 ± 0.24 0.91 27.45 ± 0.28 1.01 26.93 ± 0.03 0.11 24.90 ± 0.23 0.92 23.13 ± 0.16 0.67
27.95 ± 0.260.9326.34 ± 0.331.2428.63 ± 0.301.0626.14 ± 0.100.3829.54 ± 0.250.86
1022.0 10230.42 ± 0.08 0.26    2/3**30.19 ± 0.60 1.98   0/3**   0/3**
31.98 ± 0.170.5329.77 ± 0.391.3231.74 ± 0.792.5028.93 ± 0.080.2831.07 ± 0.371.18
1012.0 10133.24 ± 0.29 0.87    1/3**33.27 ± 1.53 4.60   0/3**   0/3**
34.34 ± 1.032.9932.84 ± 0.441.33  2/3**29.91 ± 0.100.33   0/3**
12  0/3**   1/3**  0/3**  0/3**   0/3**
  0/3**   0/3**  2/3**  0/3**   0/3**

Efficiency  98%    96%   100%   94%    104%
  97%   94%  95%  100%   87%

  0.998    0.999   0.999   0.998    0.999
  0.996   0.998  0.997  0.996   0.998

y-intercept  36.89    37.67   36.99   35.29    32.78
  38.15   36.80  38.93  38.93   40.74

CV (%) = standard deviation/mean Ct × 100.
CV was not calculated when some of the parallels failed to amplify.
: number of positives/number of experiments.
Average Ct values: mean ± standard deviation.
All PCR comparisons were run on the Rotor Gene Q MDx thermocycler (in italic font) and on the Mx3005P thermocycler (in bold).