Validation and Development of a New Automatic Algorithm for Time-Resolved Segmentation of the Left Ventricle in Magnetic Resonance Imaging

Table 5

Image processing error measurements in second observer subset. Image processing error measurements in the second observer subset () as dice similarity coefficient (DSC) and point to curve (P2C) for second observer manual delineation versus manual reference delineation and for automatic segmentation versus manual reference delineation.

Automatic segmentation versus manual reference Second observer versus manual reference
DSCP2CDSCP2C

Endocardium overall0.91 ± 0.022.0 ± 0.4 mm0.95 ± 0.011.2 ± 0.2 mm
Endocardium ED0.93 ± 0.021.8 ± 0.5 mm0.96 ± 0.011.1 ± 0.3 mm
Endocardium ES0.85 ± 0.042.4 ± 0.5 mm0.92 ± 0.031.4 ± 0.3 mm
Epicardium overall0.93 ± 0.012.2 ± 0.4 mm0.96 ± 0.011.2 ± 0.33 mm
Epicardium ED0.94 ± 0.012.0 ± 0.5 mm0.97 ± 0.011.1 ± 0.4 mm
Epicardium ES0.91 ± 0.022.4 ± 0.6 mm0.95 ± 0.011.4 ± 0.4 mm

Difference are expressed as mean ± SD. For a perfect overlap between the regions DSC should be 1 and P2C should be 0. ED = end diastole, ES = end systole.