Comparison of Imaging Characteristics of 124I PET for Determination of Optimal Energy Window on the Siemens Inveon PET

Table 3

Recovery coefficients and %SD values for 5 different rods.

Rod diameter (mm)2345
RC%SDRC%SDRC%SDRC%SD

FBP
 350~5500.2332.900.3644.090.5350.200.6465.26
 350~6000.2429.890.3940.540.5046.350.6263.38
 350~6250.2733.430.4045.780.5247.420.6262.90
 350~6500.2731.060.3842.240.4946.740.6362.11
 350~6750.2227.410.3842.260.6045.780.6061.49
 350~7000.2528.950.4041.270.5746.760.6761.51
 350~7250.2630.600.3740.650.5247.110.6562.01
 350~7500.2731.200.4144.300.5446.570.6761.64
 350~7750.2429.340.4141.950.5747.070.6562.14
 350~8000.2325.710.4141.560.5546.940.6862.98
 400~5900.2432.660.3743.360.4949.370.6666.94

For 124I PET, RC could not be calculated for a 1 mm rod size, because this rod was not discernible by 124I PET.
RC: recovery coefficient.
%SD: percent standard deviation.