Comparison of Imaging Characteristics of 124I PET for Determination of Optimal Energy Window on the Siemens Inveon PET

Table 4

Spillover ratio in air and water compartments with attenuation, scatter, and single gamma photon correction applied.

Energy windowAirWaterwSOR

350~550−5.031.011.30
350~600−4.331.261.43
350~625−3.401.801.86
350~650−2.682.872.86
350~675−4.142.152.22
350~700−6.470.791.32
350~725−5.850.391.03
350~750−5.540.260.95
350~775−5.771.581.83
350~800−5.412.362.49
400~590−5.751.401.67

Mean difference4.286.74N/A
Mean difference7.6612.79N/A

Mean difference: mean difference of spillover ratio between attenuation correction and attenuation and scatter correction.
Mean difference: mean difference of spillover ratio between attenuation correction and attenuation and scatter and single gamma photon correction.
Weighted SOR (wSOR).
Not applicable.