Research Article

Evaluation of RealStar® Alpha Herpesvirus PCR Kit for Detection of HSV-1, HSV-2, and VZV in Clinical Specimens

Table 5

Imprecision testing of the RealStar® and in-house multiplex real-time PCR assays using HSV-1, HSV-2, and VZV extracts.

DNA concentration (copies/reaction)Intra-assayInterassay
No. of positive replicatesMean Cp ± SD (% coefficient of variation)Mean Cp ± SD (% coefficient of variation)

RealStar® alpha Herpesvirus PCR assay
 HSV-1
  6.04 × 102331.41 ± 0.12 (0.38)31.38 ± 0.26 (0.84)
  6.04 × 101333.99 ± 0.32 (0.93)34.16 ± 0.36 (1.06)
 HSV-2
  1.76 × 103330.90 ± 0.08 (0.26)30.71 ± 0.26 (0.80)
  1.76 × 102332.98 ± 0.08 (0.25)33.37 ± 0.44 (1.30)
 VZV
  2.29 × 103331.54 ± 0.13 (0.41)31.84 ± 0.35 (1.11)
  2.29 × 102334.42 ± 0.39 (1.12)34.91 ± 0.68 (1.94)

In-house multiplex PCR
 HSV-1
  6.04 × 103330.79 ± 0.12 (0.39)30.80 ± 0.10 (0.33)
  6.04 × 102334.09 ± 0.15 (0.45)34.09 ± 0.18 (0.52)
  6.04 × 101337.46 ± 0.85 (2.28)37.84 ± 0.85 (2.26)
 HSV-2
  1.76 × 104330.57 ± 0.16 (0.52)30.39 ± 0.23 (0.76)
  1.76 × 103334.65 ± 0.21 (0.61)34.20 ± 0.51 (1.50)
  1.76 × 102337.79 ± 0.57 (1.52)37.63 ± 0.56 (1.50)
 VZV
  2.29 × 104330.64 ± 0.06 (0.18)30.11 ± 0.58 (1.91)
  2.29 × 103333.87 ± 0.36 (1.05)33.14 ± 0.83 (2.50)
  2.29 × 102336.09 ± 0.45 (1.25)36.13 ± 0.40 (1.11)