A Novel Method (CiMON) for Continuous Intra-Abdominal Pressure Monitoring: Pilot Test in a Pig Model

Table 1

Statistics of comparison between different device readings in the supine position.

DeviceBias (mmHg)Precision (mmHg)Limits of agreement (mmHg)Percentage error

IAPCiM−IAPspie0.20.9 9.6
IAPCiM−IAPdir2.12.5 31.6
IAPCiM−IAPivp1.52.5 24.9

Mean difference between different device readings (bias), precision (standard deviation), percentage error (%), and limits of agreement in the supine position. CiMON balloon-tipped catheter pressure (IAPCiM), Spiegelberg balloon-tipped catheter pressure (IAPspie), direct intraperitoneal pressure (IAPdir), and intravesical pressure (IAPivp) measurement.