Research Article

Development and Validation of Stability-Indicating GC-FID Method for the Quantitation of Memantine Hydrochloride and Its Nonchromophoric Impurities in Bulk and Pharmaceutical Dosages

Table 5

Accuracy data of related substances.

ComponentRecovery results (mean % recovery (%RSD))a
LOQ level; amount (% w/w)50% of specification levelb; amount (% w/w)100% of specification levelb; amount (% w/w)150% of specification levelb; amount (% w/w)

Imp-A116.67 (0.00)106.67 (3.00)104.00 (2.00)104.15 (3.00)
Imp-B116.67 (0.00)114.86 (3.00)111.94 (2.00)113.36 (3.00)
Imp-C112.50 (0.00)102.67 (3.00)100.22 (2.00)100.30 (3.00)

aAverage % recovery and % RSD of results of three determinations.
b0.15% of all related substances.