Research Article

Validated Stability Indicating RP-HPLC Method for Simultaneous Estimation of Codeine Phosphate and Chlorpheniramine Maleate from Their Combined Liquid Dosage Form

Table 8

System suitability parameters.

S. no.ParameterCPCPM

1Retention time (min)3.479.45
2Peak asymmetry1.01.0
3Theoretical plates1110017692
4Resolution3.057.78
5Limit of detection (LOD), g/mL 2.2630.756
6Limit of quantification (LOQ), g/mL 6.8592.293