Fluoroscopy-Guided Resolution of Ingested Thrombus Leading to Functional Disturbance of a Continuous-Flow Left Ventricular Assist Device
Table 1
HVAD parameters and laboratory data over the time (anticoagulation and hemolysis).
| ā | Speed (RPM) | Flow (LPM) | Power (Watts) | INR | PTT (sec) | Total Bilirubin (mg/dL) | LDH (U/L) |
| Prior thrombus | 2600 | 5.4 | 4 | 2.2 | 50 | 0.98 | 367 | Admission | 2600 | up to 12.0 | up to 15.0 | 1.3 | 45 | 3.53 | 4478 | Heparinization | 2500 | up to 9.0 | 5.0ā10.0 | 1.7 | 117 | 4.49 | 6420 | LV-lysis | 2500 | 4.0 | 4.0 | 1.5 | 139 | 3.56 | 5676 | Discharge | 2500 | 4.5 | 4.5 | 2.2 | 41 | 0.77 | 1194 | Followup | 2600 | 5.0 | 5.0 | 2.3 | 49 | 0.63 | 378 |
|
|
INR: international normalized ratio; PTT: partial thromboplastin time; LDH: lactate dehydrogenase.
|