Research Article

Validation of Quantitative HPLC Method for Bacosides in KeenMind

Table 7

Summary of validation results.

TestLimitsConclusions/results

SpecificityNo interfering peaks with that of the target.Complies, however, bacoside peaks are coeluting

Linearity
(calibration coefficient)
The Y-intercept should not be more than ±2%.0.38%
Linearity

Instrument precisionThe % RSD of the retention time is ≤1.0%0.24% RSD
The % RSD of the peak area is ≤10.0%0.073% RSD

Detection limit ≥ 30.00445 mg/mL

Quantitation limit ≥ 100.0089 mg/mL

Method precisionThe mean result at method working strength is within the specificationPass (1.02 %w/w)
The % RSD is ≤10.0%3.0% RSD

Intermediate precisionThe mean result at method working strength is within the specificationPass (1.02 %w/w)
The % RSD is ≤10.0%2.5% RSD

Extraction efficiency>95% recovery from first extract97.7% of total recovered

Accuracy/recoveryAt concentrations ranging from 80 to 120% of nominal stated content, the recovery Bacopaside II is 90.0% to 110.0% Sample %
80100.51
100102.76
120104.97

RangePrecision, accuracy, and linearity must meet their criteria from LOQ% to 150% of the label claim0.00445–0.89 mg/mL