Research Article
Validation of Quantitative HPLC Method for Bacosides in KeenMind
Table 7
Summary of validation results.
| Test | Limits | Conclusions/results |
| Specificity | No interfering peaks with that of the target. | Complies, however, bacoside peaks are coeluting | |
| Linearity (calibration coefficient) | The Y-intercept should not be more than ±2%. | 0.38% | | Linearity | | |
| Instrument precision | The % RSD of the retention time is ≤1.0% | 0.24% RSD | | The % RSD of the peak area is ≤10.0% | 0.073% RSD | |
| Detection limit | ≥ 3 | 0.00445 mg/mL | |
| Quantitation limit | ≥ 10 | 0.0089 mg/mL | |
| Method precision | The mean result at method working strength is within the specification | Pass (1.02 %w/w) | | The % RSD is ≤10.0% | 3.0% RSD | |
| Intermediate precision | The mean result at method working strength is within the specification | Pass (1.02 %w/w) | | The % RSD is ≤10.0% | 2.5% RSD | |
| Extraction efficiency | >95% recovery from first extract | 97.7% of total recovered | |
| Accuracy/recovery | At concentrations ranging from 80 to 120% of nominal stated content, the recovery Bacopaside II is 90.0% to 110.0% | Sample | % | 80 | 100.51 | 100 | 102.76 | 120 | 104.97 |
| Range | Precision, accuracy, and linearity must meet their criteria from LOQ% to 150% of the label claim | 0.00445–0.89 mg/mL | |
|
|