Impact of Application of Multifunction Electrode (MFE) Pads on Cardiopulmonary Resuscitation Quality

Table 1

Selected parameters of chest compression in both groups.

CC parametersGroup MFE (n = 100)Group C (n = 100) value

CC rate (minute)121 ± 9122 ± 80.691
Correct CC rate (%)36 (16; 59)33 (12; 47)0.064
CC depth (mm)55 ± 552 ± 4<0.001#
Correct CC depth (%)53 (45; 60)45 (32; 55)<0.001#
Correct chest recoil (%)32 (22; 48)19 (9; 38)0.001#
Correct hands position (%)100 (99; 100)100 (98; 100)0.372
CCF (%)77.0 ± 6.573.1 ± 7.1<0.001#
No-flow time (sec)6.0 ± 1.17.3 ± 1.1<0.001#

Continuous variables are expressed as means with standard deviations (with normal distribution) or medians (25th; 75th percentile). #Statistically significant differences between studied groups (). C = control; CC = chest compression; CCF = chest compression fraction; MFE = multifunction electrode.