Research Article

Determination of Chlorinity of Water without the Use of Chromate Indicator

Table 4

Comparison of the Results of Determination of Chlorinity (g/kg) from the Mohr-Knudsen Method and the Present Methods with Phosphate/Phenolphthalein as Indicator.

TrialMohr-Knudsen MethodPhosphate/Ph’pht Method (Present) Difference

117.0617.420.38
217.0017.320.32
316.8517.871.02
417.3617.610.05
517.3517.700.35

Ave.  ± Std.Dev.17.12 ± 0.2217.58 ± 0.22