Research Article
Tin Content Determination in Canned Fruits and Vegetables by Hydride Generation Inductively Coupled Plasma Optical Emission Spectrometry
Table 1
ICP-AES operating conditions.
| Instrument | Prodigy high-dispersive ICP |
| Spectrometer | High resolution echelle polychromator | Large-format programmable array detector (L-PAD) | RF generator | 40 MHz “free running” |
| Argon flow | Coolant: 18 L min−1 | Auxiliary: 0.8 L min−1 | Nebulizer: 36 psi |
| Nebulizer | Pneumatic (glass concentric) |
| Spray chamber | Glass cyclonic |
| Hydride generator | Leeman Labs. Inc. Part No. 130-1070 | Three channel peristaltic pump (0.9 mL min−1) | T connector | Reaction coil | Output power (1.3 kW) | Plasma viewing (Radial) |
| Replicates for each analysis run | 3 |
| Sample uptake delay | 50 s |
| Integration time | 40 s |