Research Article
Determination of Mercury in Ayurvedic Dietary Supplements That Are Not Rasa Shastra Using the Hydra-C Direct Mercury Analyzer
Table 6
Comparison and practical considerations for using Hydra C versus ICP-MS.
| Hydra C | ICP-MS |
| No Sample digestion (lower cost per analysis) | Need sample digestion | No Hazardous chemical or waste | Hazardous chemical or waste | All of the Hg in each sample is collected on the amalgam prior to analysis | Hg can be lost from aqueous samples during digestion | About 5 minutes/sample | About 30 min/sample | All the Hg gets to the detector, so sensitivity is good. | ICP-MS system must provide very high sensitivity to compensate for Hg low degree of ionization in the plasma | Decrease the chance of contamination | Chance of contamination during the digestion process | No carry over effect or loss of sample (no tubing required) | Significant carry over effect and loss of sample via tubing. |
|
|