Research Article

Masking Agents Evaluation for Lead Determination by Flow Injection-Hydride Generation-Atomic Fluorescence Spectrometry Technique: Effect of KI, L-Cysteine, and 1,10-Phenanthroline

Table 1

Operation parameters of AFS instrument and FIA system.

AFS parametersChemical conditions

Argon flow rate (mL min−1)400KBH41% (m/v)
Delay time (s)4K3[Fe(CN)6]1.5% (m/v)
Reading time (s)22Carrier HCl1.5% (v/v)
Load sample time (s)8HCl medium for sample and standard solutions1.5% (v/v)
Sample volume (mL)1.3
Sampling loading flow rate (mL min−1)8