Masking Agents Evaluation for Lead Determination by Flow Injection-Hydride Generation-Atomic Fluorescence Spectrometry Technique: Effect of KI, L-Cysteine, and 1,10-Phenanthroline

Table 3

Results of the effect of foreign ions and masking agents on the analytical signal of 10 µg Pb L−1.

Foreign ion concentration (µg L−1)FI without masking agentRecovery (%)FI with KIRecovery (%)FI with L-cysteineRecovery (%)FI with 1,10-phenanthrolineRecovery (%)

Arsenic
 01071 ± 41036 ± 21026 ± 21089 ± 12
 100997 ± 3931043 ± 6101995 ± 5971097 ± 7101
 200988 ± 792968 ± 593844 ± 9821016 ± 1093
 300874 ± 0.282887 ± 392681 ± 4661009 ± 893

Copper
 01071 ± 91036 ± 111026 ± 21071 ± 9
 300734 ± 569935 ± 790631 ± 5611442 ± 17135
 600842 ± 779842 ± 790627 ± 25611442 ± 25135
 1000719 ± 467719 ± 485477 ± 2461395 ± 17130

Chromium
 01057 ± 41036 ± 11950 ± 41071 ± 9
 100770 ± 973755 ± 973763 ± 380768 ± 772
 200760 ± 1272762 ± 273747 ± 1279768 ± 1672
 300933 ± 1388932 ± 190675 ± 571768 ± 2072

Iron
 01057 ± 4
 1001025 ± 1696
 2001009 ± 1395
 3001006 ± 895
 6001019 ± 1796
 10001009 ± 795

FI: fluorescence intensity.
Measurements reported as the mean ± SD ().