Research Article

Use of the Relaxometry Technique for Quantification of Paramagnetic Ions in Aqueous Solutions and a Comparison with Other Analytical Methods

Table 2

Linear regression parameters for relaxometry and spectrophotometry calibration curves.

IonSlopeIntercept

SpinlockNi2+572 ± 30.404 ± 0.0010.9998
Cu2+1491 ± 80.402 ± 0.0020.9998
Cr3+7370 ± 1580.420 ± 0.0100.9978
Mn2+31487 ± 2820.405 ± 0.0030.9993

UNMRNi2+632 ± 410.4 ± 0.10.9998
Cu2+1141 ± 810.0 ± 0.30.9996
Cr3+9570 ± 13013.4 ± 0.20.9993
Mn2+28517 ± 29410.3 ± 0.30.9992

SPNi2+5.01 ± 0.010.0004 ± 0.00030.9999
Cu2+12.45 ± 0.040.005 ± 0.0020.9999
Cr3+16.28 ± 0.290.0365 ± 0.05130.9991

Mean ± SD (); = intercept + slope[ion], where (s−1) is the inverse of and the ion concentration is given in mol L−1; = intercept + slope[ion], where is the absorbance in 810, 422, and 394 nm for Cu2+, Cr3+, and Ni2+, respectively, and the ion concentration is given in mol L−1.