Arsenic Species Analysis at Trace Level by High Performance Liquid Chromatography with Inductively Coupled Plasma Mass Spectrometry

Table 1

Working conditions of the HPLC and ICP–MS system.

ParametersValue

ICP-MS system
RF power (W)1500
Cool flow (L min−1)14.00
Auxiliary gas flow (L min−1)0.80
Nebuliser gas flow (L min−1)1.015
Collision gas flow (L min−1)He/4.09
CCT1 flow (mL min−1)5.198
Sampler depth (mm)5.0
Spray chamber temperature/°C2.70
Operation modeKED mode
Isotopes monitored75As
Dwell time (ms)200
HPLC system
ColumnHamilton PRP–X100 column (250 mm × 4.1 mm i.d., 10 μm)
Mobile phasePhase A: 1.25 mmol/L of Na2HPO4 and 11 mmol/L of KH2PO4, pH 6.116
Phase B: 2.5 mmol/L of Na2HPO4 and 22 mmol/L of KH2PO4, pH 6.057
Flow–rate (mL/min)Gradient elution of range 0.6-1.5 mL/min
HPLC elution program0-4.9 min:100% A, 4.9-5.3 min: 100% A to 100% B, 5.3-12 min:100% B
0-4.3 min: 0.6 mL/min, 4.3-4.6 min: 0.6 to 1.0 mL/min, 4.6-4.9 min: 1.0 to 1.5 mL/min, 4.9-12 min: 1.5 mL/min.
Column temperature (°C)Room temperature
Quantitative loop (μL)20 μL