Research Article
An UPLC-MS/MS Method for Determination of Osimertinib in Rat Plasma: Application to Investigating the Effect of Ginsenoside Rg3 on the Pharmacokinetics of Osimertinib
Table 3
Intraday and interday precision and accuracy of osimertinib in rat plasma.
Analytes
Spiked concentration (ng/mL)
Intraday (
n
= 6)
Interday (
n
= 18)
Mean ± SD (ng/mL)
RSD (%)
Accuracy (%)
Mean ± SD (ng/mL)
RSD (%)
Accuracy (%)
Osimertinib
1 (LLOQ)
1.02 ± 0.11
10.4
102.5
0.99 ± 0.10
10.3
98.5
3 (QC-L)
2.94 ± 0.20
6.9
98.1
2.91 ± 0.20
6.8
97.0
30 (QC-M)
31.0 ± 0.5
1.7
103.2
30.3 ± 1.6
5.2
101.1
350 (QC-H)
368 ± 11
3.1
105.1
365 ± 13
3.5
104.3