Detecting Optic Disc on Asians by Multiscale Gaussian Filtering

1Department of Electrical and Computer Engineering, University of Waterloo, Waterloo, ON, Canada N2L 3G1
2Biometric Research Center, and Department of Computing, Hong Kong Polytechnic University, Hong Kong

Received 10 February 2012; Revised 23 April 2012; Accepted 23 April 2012

Academic Editor: Qiang Li

Copyright © 2012 Bob Zhang et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

The optic disc (OD) is an important anatomical feature in retinal images, and its detection is vital for developing automated screening programs. Currently, there is no algorithm designed to automatically detect the OD in fundus images captured from Asians which are larger and have thicker vessels compared to Caucasians. In this paper, we propose such a method to complement current algorithms using two steps: OD vessel candidate detection and OD vessel candidate matching. The first step is achieved with multiscale Gaussian filtering, scale production, and double thresholding to initially extract the vessels' directional map of various thicknesses. The map is then thinned before another threshold is applied to remove pixels with low intensities. This result forms the OD vessel candidates. In the second step, a Vessels' Directional Matched Filter (VDMF) of various dimensions is applied to the candidates to be matched, and the pixel with the smallest difference designated the OD center. We tested the proposed method on a new database consisting of 402 images from a diabetic retinopathy (DR) screening programme consisting of Asians. The OD center was successfully detected with an accuracy of 99.25% (399/402).

1. Introduction

The optic disc is a vertical oval with average dimensions of 1.76 mm (horizontally) 1.92 mm (vertically), and situated 3-4 mm to the nasal side of the fovea [1]. In fundus imaging the OD is usually brighter than its surrounding area and is the convergence of the retinal blood vessel network. Detection of the OD is useful in the diagnosis of glaucoma, optic neuropathies, optic neuritis, anterior ischemic optic neuropathy or papilledema, and optic disc drusen. It can also be used as a marker to help locate fovea/macula [24], as well as decide if the image is of the left or right eye. For diabetic retinopathy, detection of the OD assists physicians identifying neovascularization of the disc (NVD) in the advanced stage of DR, proliferative diabetic retinopathy (PDR). This makes the task of automatic OD detection both relevant and necessary. A distinction needs to be made between automatic OD detection and automatic OD boundary detection. The former refers to the location of the disc center, while the latter aims to segment the OD by detecting the boundary between the retina and the nerve head. Our work is detecting the OD center.

In the literature, OD detection can be categorized into various groups. The first group uses properties of the OD [58] such as high pixel intensity and its oval shape. Morphology [9] is also used where the OD center is the center of the brightest connected object found by thresholding an intensity image. References [1012] applied template matching to locate the center. Osareh’s et al. [10] template was the average gray-level of 25 normalized images, Lalonde et al. [11] used the Hausdorff-based template, and Youssif et al. [12] employed a Vessels’ Directional Matched Filter. Two different kinds of transforms, Hough [1315] and Watershed [16], have also been applied to locate the edges of the OD and subsequently its center. Supervised learning is another group consisting of feature extraction, and classification with a Bayesian [4] or k-NN [2, 3] classifier. A geometric model was built in [17] to represent the main retinal vessels which pass the OD center. Fuzzy convergence developed by Hoover and Goldbaum [18] determined the originating vessel map convergence point near the OD center. Even though the OD features and characteristics are well defined in fundus images, the task of automatically detecting its center is still challenging.

Ethnicity affects the OD shape. A study conducted by [19] measuring disc area, cup volume, maximal cup depth, and vertical cup-disc ratio showed that Asians have higher values in these properties compared to Caucasians. This can be clearly seen in Figure 1 where (a) [20] is Caucasian while (b) is Asian, taken from our database. Further studies [2124] demonstrated that with an increase in size of the OD, blood vessels (both arteries and veins) increase in thickness, where venular widening is linked with progression of DR [24] and narrowing of the arteriolar is associated with the risk of diabetes [24]. The increase in vessel thickness caused by an enlarged OD can be seen in Figure 2, which shows the cross-sections of both thick ((a)-(b)) and thin ((c)-(d)) vessels, taken from two individuals. The dotted line is Caucasian, while the sold line is Asian. Generally speaking for thick vessels the cross-section of Asians is about 3-4 pixels wider than Caucasians. On the other hand, for thin vessels the difference is around 1-2 pixels. In order to achieve a fair assessment we examined vessels only in the OD, comparing with the same orientation and grade of DR.

fig1
Figure 1: Comparison of Caucasian retina (a) to Asian (b).
fig2
Figure 2: Thick (a) and (b); thin (c) and (d) vessel cross-sections. The solid line is Asian while the dotted comes from Caucasian.

Most of the current algorithms working on OD detection are from Western countries; therefore, we can safely assume that the fundus images collected in their database are of Caucasians. Hence, there is a need to accurately detect the OD from Asians as state-of-the-art algorithms do not take into account the increased OD size and its subsequent consequences. This paper proposes such a method of dealing with Asian OD using both vessel and intensity information [25] to complement existing algorithms. In the first step, a vessels’ directional map representing the OD vessel candidates is calculated using multiscale Gaussian filtering with scale production and double thresholding. This accounts for the thicker vessels of Asians. A VDMF template (with bilinear interpolation) is matched to each OD vessel candidate as part of the second step [12]. The pixel candidate having the least difference with the template is assigned the OD center.

The remainder of this paper is organized as follows. Section 2 describes the material used, Section 3 presents the proposed method, and experimental results are relayed in Section 4. These results are discussed in Section 5 and a conclusion is given in Section 6.

2. Material

We constructed a new database obtained from a DR screening programme in Harbin, China. The patients’ consent was obtained according to the Declaration of Helsinki and that the Ethical Committee of the Institution in which the work was performed has approved it. This database will be referred to as the HIT database. HIT consists of 402 images, broken down into 46 normal and 356 pathological (all with DR). The 356 DR images were further divided into 181 mild nonproliferative diabetic retinopathy (NPDR), and 175 moderate NPDR, classified based on [26]. The images were captured in digital form using a Canon CR-DGi Non-Mydriatic Retinal Camera at 45° field of view (FOV). The size of each image is pixels with 24 bits and in compressed JPEG format. Figure 3 shows examples of images from HIT; the ones on the left are healthy retinas while the others have moderate NPDR. The OD center in each image of the database was manually segmented by the first author.

fig3
Figure 3: (a), (c), and (e) Normal fundus image from HIT. (b), (d), and (f) Fundus diagnosed with moderate NPDR from HIT.

3. Method

As mentioned above, the proposed method based on Youssif et al. [12] consists of two steps (illustrated in Figure 4), OD vessel candidate detection and matching. The following section explains each step in more detail.

727154.fig.004
Figure 4: Steps of the proposed method shown in rectangles and its control flow given by the arrow.
3.1. OD Vessel Candidate Detection

Multiscale Gaussian filtering is based on matched filters first proposed in [27] to detect vessels. It makes use of the prior knowledge that the cross-section of vessels can be approximated by a Gaussian function. Therefore, a Gaussian-shaped filter can be used to “match” the vessels. The idea of multiscale allows more than one scale to be used which can match vessels of various widths.   The multiscale Gaussian filter is defined as where represents the scale of the filter; is used to normalize the mean value of the filter to 0 so that the smooth background can be removed after filtering; is the length of the neighborhood along the y-axis to smooth noise; is a constant and is usually set as 3 because more than 99% of the area under the Gaussian curve lies within the range of . The parameter is also chosen based on . When is small, is set relatively small, and vice versa. In the actual implementation will be rotated to detect the vessels of different orientations.

The response of multiscale Gaussian filtering can be expressed by where is a normalized green channel image and denotes convolution. The scale production is defined as the product of filter responses at two scales and :

Double thresholding is then applied to to generate a binary image where a one-pixel-wide center line of the vessel is detected using morphological thinning. The vessels’ directional map is calculated by finding the corresponding orientation that produced the maximum response with (use of the vessel feature). This map is thinned by multiplying with the center-line vessel. Using the notation that the OD has higher pixel intensities (use of the intensity feature) than its surrounding retinal background, any pixels less than 0.9 in are removed. A neighborhood of each remaining pixel is extracted in order to better represent the OD vessels. This results in the OD vessel candidates. Figures 5(a)5(e) illustrates these steps using an example. In some situations hard exudates may also be part of the OD vessel candidates since their pixel intensity is also high. However, these objects are not made of vessels and will be removed in the following step. In [12] normalization of luminosity and contrast were applied to the retinal images, with its vessels extracted using a one scale 2D Gaussian Matched Filter [27].

fig5
Figure 5: Steps of the proposed method applied to a fundus image (a). (b) is the scale production of (a). The result of double thresholding on (b) is (c). The thinned vessels of (c) is (d). (e) is the OD vessel candidates after removing pixels with low intensities. The detected OD center is illustrated in (f) by a cross.
3.2. OD Vessel Candidate Matching

We define a template as the VDMF shown in Figure 6. Each value in this template represents a different orientation (rad), , where 8 are used instead of 12 [12]. In order to account for the various sizes of vessel maps in the OD, bilinear interpolation was employed to restructure the template into and . The values and sizes are specifically tuned for HIT. Reference [12] employed 4 different template dimensions suited for their database. Each of the two templates is matched to the candidates with an absolute difference calculated. The candidate pixel with least accumulated difference is assigned the OD center. Figure 5(f) shows the final result.

727154.fig.006
Figure 6: A vessels’ directional matched filter designed for HIT.

4. Experimental Results

The key parameters in our experiments are set as follows: , and , with corresponding , and , and 8 orientations (refer to (1)). The scale production (see (3)) of and is combined along with the result of and using logical OR after double thresholding. These parameters were chosen based on our experimental experience. It took 29 secs to process each image using a 2.40 GHz Intel Centrino Pro with 2 GB RAM. In order to improve the computation time of the proposed method every image in HIT was resized by 0.5 to pixels. In the literature, the detected OD center is considered correct if it is positioned within 60 pixels of the manually identified center [12, 17, 18]. However, the images [12, 17, 18] used were pixels.

In order to compensate for the larger images in HIT, the distance from the manually identified center to the detected OD center is increased to 80 pixels. For completeness we have included results using both 80 and 60 pixels and compared the proposed method to Youssif et al. [12] as well as single scale. The original parameters stated in Youssif et al. [12] were used. Single scale [27], as the name suggests, uses one in (1) instead of many (as is the case with multi-scale). This is combined with VDMF to detect the OD and evaluated against the proposed method. These can be found in Tables 1 and 2, respectively. Columns 2–5 in Tables 1 and 2 represent the number of normal, mild NPDR, moderate NPDR, and the total number of images. With 80 pixels the OD center in all of normal was detected. Only 1 and 2 were missed in mild and moderate NPDR which lead to an overall success rate of 99.25% (399/402). As for Youssif et al. [12], 2 were not detected in normal, 4 in mild NPDR, and 4 again in moderate NPDR. The result of single scale can be found in the last row, where 44, 178, and 170 were detected, respectively, for each of the three DR classes. Using 60 pixels (see Table 2), 1 was missed in all of normal, 4 from mild NPDR, and 4 in moderate NPDR. The success rate was 97.76% (393/402). Obviously, by raising the standard from 80 to 60 pixels the success rate is bound to drop which is the situation here. In cases where the OD center failed to be detected, one of the main reasons was low contrast exhibited by the image. We admit in such cases our method may fail. Youssif et al. [12] on the other hand failed to detect 9, 13, and 16 from the three classes, while single scale missed 8 from normal, 12 from mild NPDR, and 18 in moderate NPDR.

tab1
Table 1: OD detection result on HIT using 80 pixels as standard.
tab2
Table 2: OD detection result on HIT using 60 pixels as standard.

Table 3, which has the same format as the previous two, illustrates the average distance between the estimated and manually identified OD centers. In this table the average distance of normal is 27.9 pixels, 28.8 for mild NPDR, 31.1 in the case of moderate NPDR, and 29.3 for the average of all three. This compares to averages of 35.7 using Youssif et al. [12] and 40.1 with single scale. Figure 7 shows the OD detection results of the proposed method on images from HIT.

tab3
Table 3: Average distance (in pixels) of detected and actual OD.
fig7
Figure 7: Results of the proposed method where a white cross represents the detected OD center.

5. Discussion

The variables in (1) were chosen based on extensive experiments. Because there is no ground truth for the vessel maps, our criteria is judged on the visual result (binary image of segmented vessel map) of several images. We tested and found the current parameters in Section 4 to give the best result. As for the two template sizes in Section  3.2, corresponds to covering a smaller vessel map in the OD, while is for maps of medium-to-large dimensions. These are shown in Figure 8.

fig8
Figure 8: A small vessel map is covered with the 61 × 21 template (a). Medium-to-large maps are covered using the 121 × 41 template (b).

Figure 9 depicts the result of increasing/decreasing the template size by a factor of . In total there were 4 templates and 3 pairings, with (original), with , and with . Each pair corresponds to a point in Figure 9 which plots the average distance of detected and actual OD. The left (31.8 pixels) and right (37.7 pixels) points in Figure 9 are based on a scale factor of −1.5 and +1.5, respectively. This scale factor was chosen specifically for the HIT database. As can be seen from Figure 8 anything greater or smaller would not cover the OD efficiently affecting OD vessel candidate matching. The smallest distance was achieved with the current templates discussed above (see Section 3.2) and shown in Figure 9 as the central point.

727154.fig.009
Figure 9: The result of average distance compared with various scale factors applied to the templates. From right to left, the first point is +1.5, next is the current template size, and the left point is −1.5.

The results from Table 1 show that even if the proposed method, Youssif et al. [12] and single scale achieve similar accuracy using 80 pixels (7 image difference), their gap is much wider with 60 pixels (29 image variation). Subsequently, the average distance between actual and estimated OD centers of the proposed method and single scale is close to 11 pixels, while for Youssif et al. [12] it is slightly less than 6 pixels. This shows the proposed is more accurate at OD detection compared to the others. Note in Tables 1 and 2 Youssif et al. [12] and single scale detected the same number of OD. We believe this to be the case because both methods incorporate single-scale vessel extraction in order to locate OD vessel candidates.

Figure 10 illustrates a sample of the OD detection results of the proposed method, single scale, and Youssif et al. [12]. In total three images are used, one from each DR group. The columns in Figure 10 represent the different OD detection methods. In this figure you can clearly see that the proposed method was able to detect the OD (all marked with a white cross except (a) which used a black cross), while the other two methods failed. Each of the retinal images in this figure is composed of a vascular map that consists of both thick and thin vessels. Hence, multiscale with VDMF was able to triumph, while single scale and Youssif et al. [12] both failed, due to its inability to match both thick and thin vessels found in Asians as well as account for a larger OD. This underlines the necessity of the proposed method at OD detection on Asians.

fig10
Figure 10: Visual comparison of the OD detection results (proposed method with others).

Other advantages of the proposed method include its capacity to calculate a vessels’ directional map (in OD Vessel Candidate Detection) implicitly, while extracting the vessels without any additional algorithms as needed in [28, 29]. Also, our proposed method does not make any assumptions about the location of the OD in the image. In computer-aided diagnosis of HIT, vessel extraction is a first step. Hence, the time needed to extract the vessels can be discounted.

6. Conclusion

This paper presented a method for automatic detection of Asian OD to complement existing algorithms using two steps, OD vessel candidate detection, and OD vessel candidate matching. The study was deemed necessary since an Asian OD is typically larger than Caucasians and has thicker vessels. The proposed method makes use of the OD intensity information by removing low pixel values, and incorporates the vessel information in the form of a vessel’s directional map. The use of multiscale Gaussian filtering at extracting both thin/thick vessels and the VDMF can be attributed to the high OD detection results seen in Tables 13. These tables along with Figure 10 show that algorithms developed in the West cannot adequately deal with the added properties of an Asian OD. Also, the use of single scale and Youssif et al. [12] are insufficient to match the various widths of the vessel.

We have carefully examined the cases in which the proposed method failed to correctly detect the OD center. In the majority of failed detections, poor contrast, imaging artifacts, and the presence of pathology, or a combination of these factors have attributed to the problem. When contrast over the OD is low, this affects the result of OD vessel candidate detection. As part of the future work, the aspect of contrast enhancement will be integrated to deal with such cases.

References

  1. W. Tasman and E. Jaeger, Eds., Duane's Ophthalmology, Lippincott Williams & Wilkins, 2006.
  2. M. Niemeijer, M. D. Abramoff, and B. van Ginneken, “Automated localization of the optic disc and the fovea,” in Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBS '08), pp. 3538–3541, Vancouver, BC, Canada, August 2008. View at Scopus
  3. M. Niemeijer, M. D. Abràmoff, and B. van Ginneken, “Segmentation of the optic disc, macula and vascular arch in fundus photographs,” IEEE Transactions on Medical Imaging, vol. 26, no. 1, pp. 116–127, 2007. View at Publisher · View at Google Scholar · View at Scopus
  4. K. W. Tobin, E. Chaum, V. P. Govindasamy, and T. P. Karnowski, “Detection of anatomic structures in human retinal imagery,” IEEE Transactions on Medical Imaging, vol. 26, no. 12, pp. 1729–1739, 2007. View at Publisher · View at Google Scholar · View at Scopus
  5. C. Sinthanayothin, J. F. Boyce, H. L. Cook, and T. H. Williamson, “Automated localisation of the optic disc, fovea, and retinal blood vessels from digital colour fundus images,” British Journal of Ophthalmology, vol. 83, no. 8, pp. 902–910, 1999. View at Scopus
  6. H. Li and O. Chutatape, “Automatic location of optic disk in retinal images,” in Proceedings of the IEEE International Conference on Image Processing (ICIP '01), vol. 2, pp. 837–840, October 2001. View at Scopus
  7. H. Li and O. Chutatape, “A model-based approach for automated feature extraction in fundus images,” in Proceedings of the 9th IEEE International Conference on Computer Vision (ICCV '03), vol. 1, pp. 394–399, Nice, France, October 2003. View at Scopus
  8. R. Chrástek, M. Wolf, K. Donath, G. Michelson, and H. Niemann, “Optic disc segmentation in retinal images,” in Bildverarbeitung für die Medizin 2002, pp. 263–266, 2002.
  9. T. Walter and J. C. Klein, “Segmentation of color fundus images of the human retina: detection of the optic disc and the vascular tree using morphological techniques,” in Proceedings of the 2nd International Symposium on Medical Data Analysis (ISMDA '01), vol. 2199 of Lecture Notes in Computer Science, pp. 282–287, 2001.
  10. A. Osareh, M. Mirmehdi, B. Thomas, and R. Markham, “Comparison of colour spaces for optic disc localisation in retinal images,” in Proceedings of the 16th International Conference on Pattern Recognition (ICPR'02), vol. 1, pp. 743–746, 2002.
  11. M. Lalonde, M. Beaulieu, and L. Gagnon, “Fast and robust optic disc detection using pyramidal decomposition and Hausdorff-based template matching,” IEEE Transactions on Medical Imaging, vol. 20, no. 11, pp. 1193–1200, 2001. View at Publisher · View at Google Scholar · View at Scopus
  12. A. Youssif, A. Ghalwash, and A. Ghoneim, “Optic disc detection from normalized digital fundus images by means of a vessels' direction matched filter,” IEEE Transactions on Medical Imaging, vol. 27, no. 1, pp. 11–18, 2008. View at Publisher · View at Google Scholar · View at Scopus
  13. R. A. Abdel-Ghafar and T. Morris, “Progress towards automated detection and characterization of the optic disc in glaucoma and diabetic retinopathy,” Informatics for Health and Social Care, vol. 32, no. 1, pp. 19–25, 2007. View at Publisher · View at Google Scholar · View at Scopus
  14. S. F. Barrett, E. Naess, and T. Molvik, “Employing the Hough transform to locate the optic disk,” Biomedical Sciences Instrumentation, vol. 37, pp. 81–86, 2001. View at Scopus
  15. X. Zhu and R. M. Rangayyan, “Detection of the optic disc in images of the retina using the Hough transform,” in Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBS '08), pp. 3546–3549, August 2008. View at Scopus
  16. J. Hajer, H. Kamel, and E. Noureddine, “Localization of the optic disk in retinal image using the “watersnake”,” in Proceedings of the International Conference on Computer and Communication Engineering (ICCCE '08), pp. 947–951, Kuala Lumpur, Malaysia, May 2008. View at Publisher · View at Google Scholar · View at Scopus
  17. M. Foracchia, E. Grisan, and A. Ruggeri, “Detection of optic disc in retinal images by means of a geometrical model of vessel structure,” IEEE Transactions on Medical Imaging, vol. 23, no. 10, pp. 1189–1195, 2004. View at Publisher · View at Google Scholar · View at Scopus
  18. A. Hoover and M. Goldbaum, “Locating the optic nerve in a retinal image using the fuzzy convergence of the blood vessels,” IEEE Transactions on Medical Imaging, vol. 22, no. 8, pp. 951–958, 2003. View at Publisher · View at Google Scholar · View at Scopus
  19. C. S. Tsai, L. Zangwill, C. Gonzalez et al., “Ethnic differences in optic nerve head topography,” Journal of Glaucoma, vol. 4, no. 4, pp. 248–257, 1995. View at Scopus
  20. M. Niemeijer, B. van Ginneken, M. J. Cree et al., “Retinopathy online challenge: automatic detection of microaneurysms in digital color fundus photographs,” IEEE Transactions on Medical Imaging, vol. 29, no. 1, pp. 185–195, 2010. View at Publisher · View at Google Scholar · View at Scopus
  21. N. Cheung, S. Huynh, J. J. Wang et al., “Relationships of retinal vessel diameters with optic disc, macular and retinal nerve fiber layer parameters in 6-year-old children,” Investigative Ophthalmology and Visual Science, vol. 49, no. 6, pp. 2403–2408, 2008. View at Publisher · View at Google Scholar · View at Scopus
  22. N. Cheung, L. Tong, G. Tikellis et al., “Relationship of retinal vascular caliber with optic disc diameter in children,” Investigative Ophthalmology and Visual Science, vol. 48, no. 11, pp. 4945–4948, 2007. View at Publisher · View at Google Scholar · View at Scopus
  23. C. Samarawickrama, S. C. Huynh, J. J. Wang et al., “Relationship between retinal structures and retinal vessel caliber in normal adolescents,” Investigative Ophthalmology and Visual Science, vol. 50, no. 12, pp. 5619–5624, 2009. View at Publisher · View at Google Scholar · View at Scopus
  24. T. T. Nguyen and T. Y. Wong, “Retinal vascular changes and diabetic retinopathy,” Current Diabetes Reports, vol. 9, no. 4, pp. 277–283, 2009. View at Publisher · View at Google Scholar · View at Scopus
  25. B. Zhang and F. Karray, “Optic disc detection by multi-scale Gaussian filtering with scale production and a vessels' directional matched filter,” in Proceedings of the 2nd International Conference on Medical Biometrics, pp. 173–180, 2010.
  26. C. P. Wilkinson, F. L. Ferris, R. E. Klein et al., “Proposed international clinical diabetic retinopathy and diabetic macular edema disease severity scales,” Ophthalmology, vol. 110, no. 9, pp. 1677–1682, 2003. View at Publisher · View at Google Scholar · View at Scopus
  27. S. Chaudhuri, S. Chatterjee, N. Katz, M. Nelson, and M. Goldbaum, “Detection of blood vessels in retinal images using two-dimensional matched filters,” IEEE Transactions on Medical Imaging, vol. 8, no. 3, pp. 263–269, 1989. View at Publisher · View at Google Scholar · View at Scopus
  28. K. W. Tobin, E. Chaum, V. P. Govindasamy, T. P. Karnowski, and O. Sezer, “Characterization of the optic disc in retinal imagery using a probabilistic approach,” in International Symposium on Medical Imaging, vol. 6144 of Proceedings of the SPIE, pp. 1088–1097, San Diego, Calif, USA, February 2006. View at Publisher · View at Google Scholar · View at Scopus
  29. M. D. Abràmoff and M. Niemeijer, “The automatic detection of the optic disc location in retinal images using optic disc location regression,” in Proceedings of the 28th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBS '06), pp. 4432–4435, New York, NY, USA, September 2006. View at Publisher · View at Google Scholar · View at Scopus