Review Article
A Review of Indocyanine Green Fluorescent Imaging in Surgery
Table 10
The loss factors and corresponding attenuations [dB] of the top five loss factors in ICGA imaging.
| Loss factor | Loss | dB |
| The fluorescence of ICG in blood | 0.0027 | −25.6 | Losses in the tissue above the blood vessel | 0.2 | −6.9 | Quantum efficiency of sensor | 0.25 | −6.0 | Transmittance of the emission filter | 0.3 | −5.2 | Diffusion losses in the lambertian surface | 0.32 | −5.0 |
| Subtotal | | −48.7 |
| Other factors together | 0.62 | −1.7 |
| Total | | −50.5 |
|
|