Abstract

Supracondylar humerus fracture (SCHF) is one of the commonest elbow fractures in children. It is common injury for children with age from four to fourteen. In current study, the finite element technique is used to evaluate two techniques, namely, parallel and crossed K-wire fixation for treatment of SCHF, using K-wire fixation.

1. Introduction

The humerus bone is the upper arm bone that connects the shoulder, by articulating the humeral head with the glenoid of the scapula, to the elbow by articulating the testal humerus with the ulna and radius, as shown in Figure 1. The supracondylar region is the area in the distal part of the humerus just above the growth plate. About 87 per cent of the elbow fractures occur in the distal humerus and about 80 per cent of them occur in the supracondylar. These fractures may be in the distal of the supracondylar region and called lower type or may be in the proximal part of the supracondylar region and called upper type [1].

Accurate reduction and stable fixation are needed to avoid posthealing complication, namely, deformity and limitations of movements. These fractures most often require surgical treatment unless the bones are held in proper position. Sometimes surgeons want to know which alternative they should give preference, for the benefits of the patients. Fixation of unstable SCHF in children by two K-wires is the treatment of the choice by many surgeons (Figure 2, [2]).

These K-wires can be parallel from lateral side or crossing from medial and lateral sides.

In present work, surgeon’s treatment of SCHF using K-wire fixation (distal humerus type) is modeled, analyzed, and evaluated.

2. Research Methodology

The research role is to gather information from CT scan and X-ray developed in KKU hospital. Based on these obtained X-ray and CT scan for particular patient, a CAD solid model can be developed. The developed CAD solid model is to be coupled with the finite-element analysis to study the strength of different injury treatments adopted by orthopedic surgeons that suit that particular patient.

The solid model and finite element models were generated using the GID preprocessing program [3]. The FE computation was carried out using Tochnog FE program [4]. Tochnog is explicit-implicit FE program that can be used in the analysis of structures, thermal, elastic, or elastic-plastic engineering problems. Tochnog and GID programs run under Linux operating system.

In the beginning, a data file of the FE model was generated using GID preprocessing program then completed using a text editor. Next, the Tochnog FE module called to run the FE analysis using the developed data file, followed by visualizing the FE results using GID postprocessing program based on the output files written by the Tochnog FE module.

The X-ray and CT scan were obtained from KKU Hospital for humerus bone. The X-ray was obtained for two perpendicular projections. The X-ray image is imported as rusted image to AutoCAD program to measure the outer dimensions and to obtain the 2D geometry of CT scan at different levels. Next, the 2D drawing is imported to GID program and joined to complete the 3D solid model.

The solid model mainly consists of three main parts or three volumes, see Figures 3 and 4: top bone (volume 1), mid bone (volume 2), and bottom bone (volume 3). Figure 3 shows the solid model development for the parallel K-wire fixation treatment, while, crossing K-wire fixation solid model is shown in Figure 4.

The mid-part of the solid model is divided into two main parts: the external (cortical bone) part with wall thickness of 2 mm which is the harder and high strength bone, and internal (Cancellous bone) which has lower strength and hardness compared to cortical bone, see Figure 5.

3. Assign Boundary and Loading Conditions on the Finite Element Model

Constrain conditions along three directions are assigned on the top surfaces of humerus bone as shown in Figure 6. Three types of loading conditions were considered, normal load of 50 N along lateral -direction, normal load of 50 N along later -direction, and combined normal loads plus 1 N·M torsion load. Assigned material properties for the developed solid model are given in Table 1.

4. Finite Element Mesh Generation

Two main finite element meshes were generated for modeling parallel K-wire fixation treatment model and the other for crossing K-wire fixation model. The finite element (FE) meshes were generated using 4-node tetrahedral elements, see Figure 7. The number of nodes and elements for the two FE meshes are shown in Table 2.

Furthermore, as shown in Figure 8, four finite element models were developed to model two types of fractures at the top and bottom positions. Moreover, a friction contact surfaces were modeled in the assigned crack areas to simulate the node slipping conditions under different loads. A low friction factor of 0.1 was assigned in the crack contact areas to simulate the patient condition immediately after the surgery treatment.

5. Finite Element Results

Both peak normal principal stresses and Von-Misses stresses are reported at the two selected crack areas, see Figures 9 and 10.

Summary of the FE results for top and bottom crack positions are shown in Tables 3 and 4, respectively. Same results are also shown in charts Figures 11 and 12, respectively. The reported FE results are given for three assigned loading conditions and for two parallel and crossing K-wire fixation techniques.

Figure 11 shows the Von-Misses stresses results for top fracture area. It is clearly observed that maximum stresses on the top fracture are minimum for crossing K-wire fixation when compared with parallel K-wire fixation. These results are reported for all the three loading conditions. For example, the maximum stresses were reduced by 62% for lateral -direction loading, and by 21% for lateral -direction loading, and 20% for combined loading.

On the other hand, Figure 12 shows the Von-Misses stresses results for bottom fracture area. It is clearly observed that maximum stresses on the bottom fracture are minimum for parallel K-wire fixation treatment when compared with crossing K-wire fixation. For example, the maximum stresses were increased by 19% for lateral -direction loading but clearly decreased by 33% for lateral -direction loading and 90% for combined loading.

6. Conclusions

There are two types of treatment techniques commonly used for supracondylar humerus fracture in children. These treatments cover inserting crossing two K-wires or parallel key-wires fixation techniques. The present work showed that crossing K-wire fixation technique is beneficial for top fracture position, while parallel K-wire fixation technique is beneficial for bottom fracture position.

Acknowledgement

The authors would like to acknowledge King Abdulaziz City for Science and Technology for the financial support of the project At-28-2.