Research Article

Root ZX Electronic Foramen Locator: An Ex Vivo Study of Its Three Models’ Precision and Reproducibility

Table 2

Position of the tip of the endodontic file relative to the apex foramen in the measurements performed until 0.0.

Position⁢Root ZX⁢Root ZX II⁢Root ZX ⁢Mini
%%%

<−1.000.000.000.0
−1.0 to −0.51928.1928.1825.5
−0.5 to −0.011650.01237.51237.5
0618.8721.9825.0
0.01 to 0.0500.026.326.3
0.51 to 1.013.113.126.3
1.000.013.100.0

Negative values indicate that the endodontic file is positioned inside the root canal.