The Validity and Reliability between Automated Oscillometric Measurement of Ankle-Brachial Index and Standard Measurement by Eco-Doppler in Diabetic Patients with or without Diabetic Foot

Table 2

Accuracy of the oscillometric method compared with that of the gold standard.

AccuracySensitivitySpecificity+LR−LRKappa

Right limb95.22%94.34%95.48%20.870.0590.0000.869
Left limb96.94%96.43%97.11%33.360.0360.0000.919
Combined analysis97.37%94.50%98.29%55.120.0560.0010.928