International Journal of Electrochemistry
Volume 2012 (2012), Article ID 243031, 8 pages
http://dx.doi.org/10.1155/2012/243031
Research Article

Simultaneous Determination of Hydroquinone, Catechol and Resorcinol at Graphene Doped Carbon Ionic Liquid Electrode

Anhui key laboratory of Chemo-Biosensing, School of Chemistry and Materials Science, Anhui Normal University, Wuhu 241000, China

Received 10 March 2012; Accepted 28 March 2012

Academic Editor: Shen-Ming Chen

Copyright © 2012 Li Ma and Guang-Chao Zhao. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

A new composite electrode has been prepared with doping graphene into the paste consisting graphite and ionic liquid, n-octyl-pyridinum hexafluorophosphate (OPFP). This electrode shows an excellent electrochemical activity for the redox of hydroquinone (HQ), catechol (CC), and resorcinol (RS). In comparison with bare paste electrode, the redox peaks of three isomers of dihydroxybenzene can be obviously, simultaneously observed at graphene doping paste electrode. Under the optimized condition, the simultaneous determination of HQ, CC, and RS in their ternary mixture can be carried out with a differential pulse voltammetric technique. The peak currents are linear to the concentration of HQ, CC, and RS in the range form to , to , and to  mol L−1, respectively. The limits of detection are  mol L−1 for HQ,  mol L−1 for CC, and  M for RS, respectively.

1. Introduction

Catechol (CC), resorcinol (RS), and hydroquinone (HQ) are three isomers of dihydroxybenzene, which have been widely recognized as important environmental pollutants by the US Environmental Protection Agency (EPA) and the European Union (EU) due to their high toxicity and low degradability in the ecological environment [1]. The absorption of CC or HQ from the gastrointestinal tract can induce some disease such as renal tube degeneration and liver function decrease [2]. In addition, inhalation of high concentration of RS can directly lead to the death of human beings. Therefore, a highly sensitive and selective analytical method is necessary for the simultaneous determination of the three dihydroxybenzene isomers. Up to now, various methods have been exploited for the determination of the dihydroxybenzenes, such as high performance liquid chromatography (HPLC) [3, 4], spectrophotometry [5, 6] Electrochemiluminescence [7], synchronous fluorescence [8], gas chromatography/mass spectrometry [9], capillary electrochromatography [10], pH-based-flow injection analysis [11], and electrochemical methods [1223]. Among these methods, electrochemical method is of particular interest due to its high sensitivity, low cost, good selectivity, and rapid detection. For most electrochemical methods, more attention is focused on the simultaneous determination of CC and HQ due to their redox peaks are overlapped at ordinary electrode [1220]. However, the simultaneous determination of HQ, CC, and RS by electrochemical methods is few reported. Recently, a number of novel material-based electrochemical electrodes for detection of dihydroxybenzene isomers have been reported. For example, an enhanced electrochemical platform based on graphene-polyoxometalate nanomaterials for sensitive determination of diphenolic compounds had been reported [21] and the glassy carbon electrode (GCE) modified by multiwall carbon nanotubes (MWCNTs) can separate oxidation peaks of HQ and CC [12]. Moreover, the simultaneous determinations of HQ, CC, and RS have been performed at single-wall carbon nanotube (SWCNT) [22], MWCNT [23], graphene [24], graphene-chitosan [25], graphene/BMIMPF6 composites [26], and polymer functionalized graphene [27] modified GCE. Nevertheless, it is still interesting to investigate novel electrode material for the simultaneous determination of HQ, CC, and RS.

The use of solid materials, such as composite electrodes, as detectors for electrochemical analysis is commonly known [28]. Over the past five decades, carbon paste electrode as a type of composite electrode that was composed by a mixture of carbon powder and a binder has become one of the most popular electrode used as the sensors and detectors [29]. Generally, nonconductive reagents, such as mineral oil or paraffin oil were used as binders, but their electrochemical capability was poor due to their low conductivity. Therefore, it is still interesting to find a better binder to promote the performance of the carbon paste electrode. In the past years, ionic liquid (IL) has been proved to be an efficient binders and/or modifiers in the preparation of carbon composite electrodes [30]. By substituting the nonconductive organic binder with IL, carbon ionic liquid electrode (CILE) has been exhibited the advantages such as high conductivity, ease of fabrication, low cost, and renewable surface [31]. Maleki and coworkers fabricated n-octyl-pyridinium hexafluorophosphate-modified carbon paste electrode for the simultaneously determination of the dopamine, ascorbic acid, and uric acid [32]. Shang et al. reported that the direct electrochemistry of glucose oxidase (GOx) had been achieved by its direct immobilization on CILE with a conductive hydrophobic ionic liquid, 1-butyl pyridinium hexafluorophosphate as binder [33], which demonstrated that CILE has its unique superiority.

In recent years, graphene, considered as a “rising star” nanostructured carbon material, is a counterpart of graphite with well-separated two-dimensional aromatic sheets composed of sp2-bonded carbon atoms. By virtue of its unique characteristic such as high surface area [34], remarkable electrical conductivity [35], and high thermal conductivity [36], graphene nanosheets have attracted more and more attentions in the preparation of sensors and biosensors. It has been reported that graphene was used as an electrode material for sensing a range of target analytes such as nitric oxide [37], dopamine [38], and hydrogen peroxide [39]. Recently, a graphene-doped carbon paste electrode was prepared with the addition of graphene into the carbon paste mixture and used for the successful determination of ascorbic acid [40]. Therefore, graphene should be an excellent modifier to fabricate carbon paste electrode.

In this paper, the graphene-doped carbon ionic liquid electrode was prepared, and the fabricated electrode can respond to all three isomers of dihydroxybenzene. Based on this, a simple and fast assay for the simultaneous determination of HQ, CC, and RS can be constructed with differential pulse voltammetry. Meanwhile, the presented method has been applied successively to simultaneous determination of HQ, CC, and RS in wastewater samples with good recovery.

2. Experimental

2.1. Reagents

Catechol, resorcinol, and hydroquinone were purchased from Shanghai Chemical Reagent Corporation (Shanghai, China). Graphene were purchased from Nanjing JCNANO Tech Co., Ltd. The ionic liquids, octyl-pyridinum hexafluorophosphate (OPFP), were purchased from Chengjie Chemical Co., Ltd. (Shanghai, China). Graphite powder (average particle size 30 μm) and other chemicals are of analytical grade. All solution was prepared by using doubly distilled water.

2.2. Apparatus

All the electrochemical experiments were carried out in a three-electrode cell controlled by CHI 660B electrochemical workstation (CH Instruments, USA). Graphene-doped CILE was used as the working electrode. An Ag/AgCl electrode and a platinum wire electrode were used as the reference and the auxiliary electrodes, respectively. All the measurements were carried out at room temperature.

2.3. Preparation of Graphene-Doped CILE

The bare CILE was prepared by hand-mixing, in a mortar, the graphite powder and OPFP with a ratio of 50/50 (w/w). A portion of the resulting paste was packed firmly into one end of a glass tube (Φ = 3 mm), and a copper wire was inserted through the opposite end to establish an electrical contact. The electrode was then heated either in an oven to a temperature higher than the melting point of OPFP (mp 65°C) or simply by using a hair drier for 2 min. It was then left to cool to room temperature. A new surface was obtained by smoothing the electrode onto a weighing paper. The graphene-doped CILE was fabricated using the same procedure except the introduction of a portion of graphene into the bare CILE.

3. Results and Discussion

3.1. Electrochemical Characteristics of the Graphene-Doped CILE

Generally, is used as a redox probe to characterize new prepared electrodes. In order to investigate the action of graphene doped into CILE, cyclic voltammetric experiments were carried out in 0.1 M KCl solution containing 0.1 mM . Figure 1 shows the typical cyclic voltammograms (CVs) recorded at the graphene-doped CILE and CILE, respectively. It is obvious that the electrochemical response of at graphene-doped CILE is improved, which attributes to the excellent conducting property of graphene. Meanwhile, it also indicated that the graphene was doped into the CILE successfully.

243031.fig.001
Figure 1: CVs for 0.1 M KCl solution containing 0.1 mM at CILE (a) and graphene-doped CILE (b), scan rate: 100 mV/s.
3.2. Electrochemical Responses of HQ, CC, and RS at the Doped CILE

In order to investigate the electrochemical response of the doped electrode to three isomers of dihydroxybenzene, CV experiments were carried out in acetate buffer (HAc-NaAc) solution containing HQ, CC, or RS, respectively. Figure 2 shows the typical cyclic voltammograms of HQ (A), CC (B), and RS (C) at the bare CILE and the doped electrode, respectively. At bare CILE, the electrochemical response of three isomers of dihydroxybenzene can be observed. But the redox peak currents are very small. Especially, for 1.0 × 10−4 M RS, almost no anodic peak can be observed. As expected, under the same concentration, higher redox peak currents can be obtained at the graphene-doped electrode, which indicates the presence of graphene can accelerate the electron transfer between hydroquinone and electrode and the doped graphene provide a remarkable synergistic promotion for the electrochemistry of the dihydroxybenzene.

fig2
Figure 2: CVs of 5.0 × 10−5 M HQ (A), 5.0 × 10−5 M CC (B), and 1.0 × 10−4 M RS (C) in 0.10 M HAc-NaAc (pH 5.0) at CILE (a) and graphene-doped CILE (b). Scan rate: 100 mV/s.

Figure 3 showed the cyclic voltammograms of the mixture of 1 × 10−4 M HQ, CC, and RS at bare CILE and doped CILE in 0.10 M acetate buffer (pH 5.0). At the bare CILE (curve a), two abroad, unobvious anodic peaks appear at 0.322 V and 0.222 V, respectively. This is attributed to the oxidation of CC and HQ. Meanwhile, oxidation peak of RS almost cannot be observed. At the reverse scan, there are also two broad cathodic peaks appeared at about 0.259 V and 0.160 V, which correspond to the reduction peak of CC and HQ. Comparatively, at the doped CILE (curve b), three well-defined anodic peaks are observed at 0.234 V, 0.344 V, and 0.755 V for HQ, CC, and RS, respectively. In addition, two obvious cathodic peaks are obtained at 0.158 and 0.269 V, corresponding to the reduction of the oxidation products of HQ and CC, respectively. It is clear that the anodic or cathodic peaks of CC, RS, and HQ can be distinguished effectively and all peak currents can be detected independently at the doped CILE. It is no doubt that the increase in peak currents and the decrease in the separation of the peak potential at the graphene-doped CILE can be attributed to the high surface area, excellent catalytic activity, and good conductivity of graphene. This also suggests that the simultaneous determination of three isomers of dihydroxybenzene can be performed at the graphene-doped CILE.

243031.fig.003
Figure 3: CVs of the mixture solution of 1 × 10−4 M CC, RS and HQ at CILE (a) and the graphene-doped CILE (b) in 0.10 M acetate buffer (pH 5.0). Scan rate: 100 mV/s.
3.3. Electrochemical Behaviors of HQ, CC, and RS at the Doped CILE

The effect of pH of supporting electrolyte on the electrochemical behaviors of three isomers of dihydroxybenzene was carefully investigated in the pH value range from 3.0 to 7.0 and the results were shown in Figure 4. Figure 4(a) shows the relationship between the peak potential and pH. With increasing pH, both anodic and cathodic peak potentials decrease linearly. The linear regression equations of the three isomers of dihydroxybenzene are listed in Table 1. The slopes of the five regression equations are close to the theory value of 58.5 mVpH−1 for the electrode reaction process of equal electron and proton [4244], indicating that the electrode process of  HQ, CC, and RS at graphene-doped CILE should be a two electrons and two protons process. Figure 4(b) shows the dependence of anodic peak currents of three isomers of dihydroxybenzene on pH. From Figure 4(b), it can be seen that the peak currents of HQ, CC, and RS increase with increasing pH value until pH reaches 5.0, and then they decrease when pH increases further. Therefore, considering the determination sensitivity, pH 5.0 is chosen as the optimal experimental condition.

tab1
Table 1: Linear regression equations for peal potentials of the three isomers of dihydroxybenzene versus pH.
fig4
Figure 4: The dependences of all peak potentials (a) and anodic peak currents (b) of CC, RS, and HQ on pH.

The effect of scan rates on the CV behaviors of the three dihydroxybenzene isomers at the doped CILE is also investigated. With increasing scan rate from 50 to 500 mV/s, all the redox peak currents of three dihydroxybenzene isomers increase linearly. The linear regression equations of peak currents ( or ) versus scan rate (ν) are listed in Table 2. The good linear relationship between peak currents and scan rate indicates that the electrode process of all three dihydroxybenzene isomers is a typical adsorption-controlled process.

tab2
Table 2: The linear regression equations for peak currents ( or ) of HQ, CC, and RS versus scan rate (ν).
3.4. Simultaneous Determination of HQ, CC, and RS

Differential pulse voltammetry (DPV) is performed to investigate the relationship between the peak current and concentration of three dihydroxybenzene isomers due to its higher sensitivity. The individual determination of CC, RS, or HQ in their mixtures is carried out when the concentration of one species changes whereas those of other two species remain constant. The typical cyclic voltammograms are shown Figure 5. From Figure 5(a) to 5(c), it can be seen that the determination of individual CC, HQ, or RS is independent of others. The good linear relationship between peak currents and the concentration of three dihydroxybenzene isomers can be obtained and is shown as inset in Figure 5. Under the optimal conditions, peak currents are linear to the concentrations in the range from 1.0 × 10−5 M to 4.0 × 10−4 M for HQ, from 1.0 × 10−5 M to 3.0 × 10−4 M for CC, and from 1.0 × 10−6 to 1.7 × 10−4 M for RS, respectively. The limit of detection (LOD) is 1.8 × 10−6 M for HQ, 7.4 × 10−7 M for CC, and 3.6 × 10−7 M for RS, respectively. Above results suggest that this prepared electrode allows us to simultaneously and sensitively determine of CC, RS, and HQ without interference with each other. A comparison of the proposed method with other electrochemical methods is listed in Table 3. It can be seen from the table that the proposed method shows rational linear range and acceptable LOD.

tab3
Table 3: Performance comparison of the fabricated electrode for HQ, CC, and RS detection with other electrodes.
fig5
Figure 5: Dependences of peak currents on concentrations . Inset: linear plots of versus . (a) DPVs of the doped CILE in the presence of 5.0 × 10−5 M CC and RS containing different concentrations of HQ (a–i: 1.0 × 10−5, 2.0 × 10−5, 3.0 × 10−5, 4.0 × 10−5, 5.0 × 10−5, 1.0 × 10−4, 2.0 × 10−4, 3.0 × 10−4, 4.0 × 10−4 M), (b) DPVs of the doped CILE in the presence of 5.0 × 10−5 M HQ and RS containing different concentrations of CC (a–h: 1.0 × 10−5, 2.0 × 10−5, 3.0 × 10−5, 4.0 × 10−5, 5.0 × 10−5, 1.0 × 10−4, 2.0 × 10−4, 3.0 × 10−4 M), (c) DPVs of the doped CILE in the presence of 5.0 × 10−5 M HQ and CC containing different concentrations of RS (a–i: 1.0 × 10−6, 5.0 × 10−6, 1.0 × 10−5, 2.0 × 10−5, 3.0 × 10−5, 4.0 × 10−5, 5.0 × 10−5, 1.0 × 10−4, 1.7 × 10−4 M).
3.5. Reproducibility and Stability of Graphene-Doped CILE

The reproducibility for ten doped CILEs was investigated by comparing the DPV peak current of 5 × 10−5 M HQ, CC, and RS in a mixed solution. The relative standard deviation (RSD) is 2.53% for HQ, 2.86% for CC, and 3.70% for RS, indicating an acceptable reproducibility. Additionally, the stability of the doped electrode is also tested. The results show that the peak currents remained 94% of their initial values after the electrode was kept for 15 days.

3.6. Sample Analysis

In order to assess the possible applications of the proposed method, the simultaneous determinations of HQ, CC, and RS in local tap water samples and lake water samples were performed. However, no aimed analytes can be detected. Then, 10 μL of different standard concentrations of the dihydroxybenzenes was added into 5.0 mL water sample. The amounts of HQ, CC, and RS in water samples were then determined again by calibration method, and the results are listed in Table 4. The recoveries were in the range from 98.0% to 106.0%. Therefore, the simultaneous determination of HQ, CC, and RS with the prepared electrode is feasible.

tab4
Table 4: Recovery results for HQ, CC, and RS in water samples.

4. Conclusions

In this paper, a simple and sensitive electrochemical analytical method for simultaneous determination of HQ, CC, and RS has been developed employing the graphene-doped carbon ionic liquid electrode. The electrode combines the unique and attractive electrocatalytic behavior of graphene and the excellent reproducibility and good long stability of IL. The proposed electrode can be further used for multicomponent analysis in environmental control and chemical industry.

References

  1. T. Xie, Q. Liu, Y. Shi, and Q. Liu, “Simultaneous determination of positional isomers of benzenediols by capillary zone electrophoresis with square wave amperometric detection,” Journal of Chromatography A, vol. 1109, no. 2, pp. 317–321, 2006. View at Publisher · View at Google Scholar · View at Scopus
  2. M. Lv, M. Wei, F. Rong, C. Terashima, A. Fujishima, and Z. Z. Gu, “Electrochemical detection of catechol based on as-grown and nanograss array boron-doped diamond electrodes,” Electroanalysis, vol. 22, no. 2, pp. 199–203, 2010. View at Publisher · View at Google Scholar · View at Scopus
  3. L.-H. Wang and Y.-P. Kuo, “Simultaneous quantitative determination of resorcinol and 1-naphthol in haircolor products by high-performance liquid chromatography,” Chromatographia, vol. 49, no. 3-4, pp. 208–211, 1999. View at Scopus
  4. C. H. Lin, J. Y. Sheu, H. L. Wu, and Y. L. Huang, “Determination of hydroquinone in cosmetic emulsion using microdialysis sampling coupled with high-performance liquid chromatography,” Journal of Pharmaceutical and Biomedical Analysis, vol. 38, no. 3, pp. 414–419, 2005.
  5. P. Nagaraja, R. A. Vasantha, and K. R. Sunitha, “A new sensitive and selective spectrophotometric method for the determination of catechol derivatives and its pharmaceutical preparations,” Journal of Pharmaceutical and Biomedical Analysis, vol. 25, no. 3-4, pp. 417–424, 2001. View at Publisher · View at Google Scholar · View at Scopus
  6. P. Nagaraja, R. A. Vasantha, and K. R. Sunitha, “A sensitive and selective spectrophotometric estimation of catechol derivatives in pharmaceutical preparations,” Talanta, vol. 55, no. 6, pp. 1039–1046, 2001. View at Publisher · View at Google Scholar · View at Scopus
  7. S. F. Li, X. Z. Li, J. Xu, and X. W. Wei, “Flow-injection chemiluminescence determination of polyphenols using luminol-NaIO4-gold nanoparticles system,” Talanta, vol. 75, no. 1, pp. 32–37, 2008. View at Publisher · View at Google Scholar · View at Scopus
  8. M. F. Pistonesi, M. S. Di Nezio, M. E. Centurión, M. E. Palomeque, A. G. Lista, and B. S. Fernández Band, “Determination of phenol, resorcinol and hydroquinone in air samples by synchronous fluorescence using partial least-squares (PLS),” Talanta, vol. 69, no. 5, pp. 1265–1268, 2006. View at Publisher · View at Google Scholar · View at Scopus
  9. S. C. Moldoveanu and M. Kiser, “Gas chromatography/mass spectrometry versus liquid chromatography/fluorescence detection in the analysis of phenols in mainstream cigarette smoke,” Journal of Chromatography A, vol. 1141, no. 1, pp. 90–97, 2007. View at Publisher · View at Google Scholar · View at Scopus
  10. N. Guan, Z. Zeng, Y. Wang, E. Fu, and J. Cheng, “Open tubular capillary electrochromatography in fused-silica capillaries chemically bonded with macrocyclic dioxopolyamine,” Analytica Chimica Acta, vol. 418, no. 2, pp. 145–151, 2000. View at Publisher · View at Google Scholar · View at Scopus
  11. J. A. Garcia-Mesa and R. Mateos, “Direct automatic determination of bitterness and total phenolic compounds in virgin olive oil using a pH-based flow-injection analysis system,” Journal of Agricultural and Food Chemistry, vol. 55, no. 10, pp. 3863–3868, 2007. View at Publisher · View at Google Scholar · View at Scopus
  12. H. Qi and C. Zhang, “Simultaneous determination of hydroquinone and catechol at a glassy carbon electrode modified with multiwall carbon nanotubes,” Electroanalysis, vol. 17, no. 10, pp. 832–838, 2005. View at Publisher · View at Google Scholar · View at Scopus
  13. J. Peng and Z. N. Gao, “Influence of micelles on the electrochemical behaviors of catechol and hydroquinone and their simultaneous determination,” Analytical and Bioanalytical Chemistry, vol. 384, no. 7-8, pp. 1525–1532, 2006. View at Publisher · View at Google Scholar · View at Scopus
  14. M. A. Ghanem, “Electrocatalytic activity and simultaneous determination of catechol and hydroquinone at mesoporous platinum electrode,” Electrochemistry Communications, vol. 9, no. 10, pp. 2501–2506, 2007. View at Publisher · View at Google Scholar · View at Scopus
  15. Y. Zhang and J. B. Zheng, “Comparative investigation on electrochemical behavior of hydroquinone at carbon ionic liquid electrode, ionic liquid modified carbon paste electrode and carbon paste electrode,” Electrochimica Acta, vol. 52, no. 25, pp. 7210–7216, 2007. View at Publisher · View at Google Scholar · View at Scopus
  16. M. Li, F. Ni, Y. Wang et al., “Sensitive and facile determination of catechol and hydroquinone simultaneously under coexistence of resorcinol with a Zn/Al layered double hydroxide film modified glassy carbon electrode,” Electroanalysis, vol. 21, no. 13, pp. 1521–1526, 2009. View at Publisher · View at Google Scholar · View at Scopus
  17. D. M. Zhao, X. H. Zhang, L. J. Feng, L. Jia, and S. F. Wang, “Simultaneous determination of hydroquinone and catechol at PASA/MWNTs composite film modified glassy carbon electrode,” Colloids and Surfaces B, vol. 74, no. 1, pp. 317–321, 2009. View at Publisher · View at Google Scholar · View at Scopus
  18. Jingjing Yu, Wei Du, Faqiong Zhao, and Baizhao Zeng, “High sensitive simultaneous determination of catechol and hydroquinone at mesoporous carbon CMK-3 electrode in comparison with multi-walled carbon nanotubes and Vulcan XC-72 carbon electrodes,” Electrochimica Acta, vol. 54, no. 3, pp. 984–988, 2009. View at Publisher · View at Google Scholar · View at Scopus
  19. P. Yang, Q. Zhu, Y. Chen, and F. Wang, “Simultaneous determination of hydroquinone and catechol Using poly(p-aminobenzoic acid) modified glassy carbon electrode,” Journal of Applied Polymer Science, vol. 113, no. 5, pp. 2881–2886, 2009. View at Publisher · View at Google Scholar · View at Scopus
  20. A. J. Saleh Ahammad, S. Sarker, M. Aminur Rahman, and J. J. Lee, “Simultaneous determination of hydroquinone and catechol at an activated glassy carbon electrode,” Electroanalysis, vol. 22, no. 6, pp. 694–700, 2010. View at Publisher · View at Google Scholar · View at Scopus
  21. L. Cao, H. Sun, J. Li, and L. Lu, “An enhanced electrochemical platform based on graphene-polyoxometalate nanomaterials for sensitive determination of diphenolic compounds,” Analytical Methods, vol. 3, no. 7, pp. 1587–1594, 2011. View at Publisher · View at Google Scholar · View at Scopus
  22. Z. Wang, S. Li, and Q. Lv, “Simultaneous determination of dihydroxybenzene isomers at single-wall carbon nanotube electrode,” Sensors and Actuators B, vol. 127, no. 2, pp. 420–425, 2007. View at Publisher · View at Google Scholar · View at Scopus
  23. Y. P. Ding, W. L. Liu, Q. S. Wu, and X. G. Wang, “Direct simultaneous determination of dihydroxybenzene isomers at C-nanotube-modified electrodes by derivative voltammetry,” Journal of Electroanalytical Chemistry, vol. 575, no. 2, pp. 275–280, 2005. View at Publisher · View at Google Scholar · View at Scopus
  24. H. Du, J. Ye, J. Zhang, X. Huang, and C. Yu, “A voltammetric sensor based on graphene-modified electrode for simultaneous determination of catechol and hydroquinone,” Journal of Electroanalytical Chemistry, vol. 650, no. 2, pp. 209–213, 2011. View at Publisher · View at Google Scholar · View at Scopus
  25. H. Yin, Q. Zhang, Y. Zhou et al., “Electrochemical behavior of catechol, resorcinol and hydroquinone at graphene-chitosan composite film modified glassy carbon electrode and their simultaneous determination in water samples,” Electrochimica Acta, vol. 56, no. 6, pp. 2748–2753, 2011. View at Publisher · View at Google Scholar · View at Scopus
  26. Z. Liu, Z. Wang, Y. Cao, Y. Jing, and Y. Liu, “High sensitive simultaneous determination of hydroquinone and catechol based on graphene/BMIMPF6 nanocomposite modified electrode,” Sensors and Actuators B, vol. 157, no. 2, pp. 540–546, 2011. View at Publisher · View at Google Scholar · View at Scopus
  27. L. Wang, Y. Zhang, Y. Du, D. Lu, Y. Zhang, and C. Wang, “Simultaneous determination of catechol and hydroquinone based on poly (diallyldimethylammonium chloride) functionalized graphene-modified glassy carbon electrode,” Journal of Solid State Electrochemistry, vol. 16, p. 1323, 2012.
  28. M. Luque, A. Ríos, and M. Valcárcel, “Validation of PVC-graphite composite electrodes for routine analytical work,” Electroanalysis, vol. 11, no. 15, pp. 1116–1123, 1999. View at Scopus
  29. I. Švancara, K. Vytřas, K. Kalcher, A. Walcarius, and J. Wang, “Carbon paste electrodes in facts, numbers, and notes: a review on the occasion of the 50-years jubilee of carbon paste in electrochemistry and electroanalysis,” Electroanalysis, vol. 21, no. 1, pp. 7–28, 2009. View at Publisher · View at Google Scholar · View at Scopus
  30. H. Liu, P. He, Z. Li et al., “An ionic liquid-type carbon paste electrode and its polyoxometalate- modified properties,” Electrochemistry Communications, vol. 7, no. 12, pp. 1357–1363, 2005. View at Publisher · View at Google Scholar · View at Scopus
  31. Afsaneh Safavi, Norouz Maleki, Fariba Tajabadi, and Elahe Farjami, “High electrocatalytic effect of palladium nanoparticle arrays electrodeposited on carbon ionic liquid electrode,” Electrochemistry Communications, vol. 9, no. 8, pp. 1963–1968, 2007. View at Publisher · View at Google Scholar · View at Scopus
  32. N. Maleki, A. Safavi, and F. Tajabadi, “High-performance carbon composite electrode based on an ionic liquid as a binder,” Analytical Chemistry, vol. 78, no. 11, pp. 3820–3826, 2006. View at Publisher · View at Google Scholar · View at Scopus
  33. X. Shang, H. Zhang, and J. Zheng, “Direct electrochemistry of glucose oxidase immobilized on NdPO4 nanoparticles/chitosan composite film on glassy carbon electrodes and its biosensing application,” Electrochemistry Communications, vol. 10, pp. 1140–1143, 2008.
  34. M. D. Stoller, S. Park, Z. Yanwu, J. An, and R. S. Ruoff, “Graphene-Based ultracapacitors,” Nano Letters, vol. 8, no. 10, pp. 3498–3502, 2008. View at Publisher · View at Google Scholar · View at Scopus
  35. K. I. Bolotin, K. J. Sikes, Z. Jiang et al., “Ultrahigh electron mobility in suspended graphene,” Solid State Communications, vol. 146, no. 9-10, pp. 351–355, 2008. View at Publisher · View at Google Scholar · View at Scopus
  36. A. A. Balandin, S. Ghosh, W. Bao et al., “Superior thermal conductivity of single-layer graphene,” Nano Letters, vol. 8, no. 3, pp. 902–907, 2008. View at Publisher · View at Google Scholar · View at Scopus
  37. J. F. Wu, M. Q. Xu, and G. C. Zhao, “Graphene-based modified electrode for the direct electron transfer of Cytochrome c and biosensing,” Electrochemistry Communications, vol. 12, no. 1, pp. 175–177, 2010. View at Publisher · View at Google Scholar · View at Scopus
  38. Y. R. Kim, S. Bong, Y. J. Kang et al., “Electrochemical detection of dopamine in the presence of ascorbic acid using graphene modified electrodes,” Biosensors and Bioelectronics, vol. 25, no. 10, pp. 2366–2369, 2010. View at Publisher · View at Google Scholar · View at Scopus
  39. K. Zhou, Y. Zhu, X. Yang, J. Luo, C. Li, and S. Luan, “A novel hydrogen peroxide biosensor based on Au-graphene-HRP-chitosan biocomposites,” Electrochimica Acta, vol. 55, no. 9, pp. 3055–3060, 2010. View at Publisher · View at Google Scholar · View at Scopus
  40. F. Li, J. Li, Y. Feng, L. Yang, and Z. Du, “Electrochemical behavior of graphene doped carbon paste electrode and its application for sensitive determination of ascorbic acid,” Sensors and Actuators B, vol. 157, no. 1, pp. 110–114, 2011. View at Publisher · View at Google Scholar · View at Scopus
  41. D. Zhang, Y. Peng, H. Qi, Q. Gao, and C. Zhang, “Application of multielectrode array modified with carbon nanotubes to simultaneous amperometric determination of dihydroxybenzene isomers,” Sensors and Actuators B, vol. 136, no. 1, pp. 113–121, 2009. View at Publisher · View at Google Scholar · View at Scopus
  42. W. Sun, M. Xi, L. Zhang, T. Zhan, H. Gao, and K. Jiao, “Electrochemical behaviors of thymine on a new ionic liquid modified carbon electrode and its detection,” Electrochimica Acta, vol. 56, no. 1, pp. 222–226, 2010. View at Publisher · View at Google Scholar · View at Scopus
  43. P. Wu, Q. Shao, Y. Hu et al., “Direct electrochemistry of glucose oxidase assembled on graphene and application to glucose detection,” Electrochimica Acta, vol. 55, no. 28, pp. 8606–8614, 2010. View at Publisher · View at Google Scholar · View at Scopus
  44. B. J. Sanghavi and A. K. Srivastava, “Simultaneous voltammetric determination of acetaminophen, aspirin and caffeine using an in situ surfactant-modified multiwalled carbon nanotube paste electrode,” Electrochimica Acta, vol. 55, no. 28, pp. 8638–8648, 2010. View at Publisher · View at Google Scholar · View at Scopus